5 Set-up and connection
Pos: 93 / 200 Christ /36 0 GT-B A La bor -Pil ot (S TANDA RDMOD ULE) /05 0 Aufs tell ung und A nschlus s/ 050-0 010 A ufst ell ort (Al pha -Delt a) @ 25\ mod _14 049 837381 97_ 68.d ocx @ 183 509 @ 2 @ 1
5.1 Installation site
Operate the freeze-dryer only in closed and dry rooms.
• The table must be stable and have a solid, even tabletop.
• Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will
be impaired.
• Keep a safety distance of at least 30 cm from the wall so that the vents
in the unit remain fully effective.
• The ambient temperature must be in the range of +10°C to +25°C.
• Do not subject the freeze-dryer to thermal stress, e.g. by positioning it
near heat generators.
• Avoid direct sunlight (UV radiation).
Pos: 94 / 010 Uni ve rsalm odul e/Leerz eil e @ 0\mo d_1 202116 244 500 _0.doc x @ 114 @ @ 1
Pos: 95 / 200 C hrist /36 0 GT-B A La bor -Pilot (S TAN DAR DMODULE ) /050 A ufstell ung und Anschl us s/05 0-0020 Ene rgiev erso rgu ng--- --- ----- ----- --- ----- ----- ----- --- -- @ 25\ mo d_14049 837 400 08_ 68.doc x @ 183 537 @ 2 @ 1
5.2 Power supply
Pos: 96 / 200 Chris t /36 1 GT-B A La bor -Pil ot (PRO JEK TE )/Al pha 1- 2 LDpl us/0 50 A ufs tell ung un d A nschl uss /0 50-002 0-0 010 Ansc hl ussar t Al pha 1- 2 LDpl us @ 31\ mod _140 808 408 0365_6 8.doc x @ 2 093 98 @ 3 @ 1
5.2.1 Connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage!
CHRIST freeze-dryers are units of safety class I.
Alpha 1-2 LDplus units have a three-wire power cord with an IEC C13
connector (see chapter 10 - "Technical data").
An equipotential bonding screw is located on the back below the mains
power input (see chapter 2.1.1 - "Functional and operating elements"). This
equipotential bonding screw can be used to perform an earth conductor
check.
Pos: 97 / 010 Uni ve rsalm odul e/Leerz eil e @ 0\mo d_1 202116 244 500 _0.doc x @ 114 @ @ 1
Pos: 98 / 200 Chris t/ 360 GT -B A La bor -Pil ot (S TAND ARD MODULE ) /05 0 Aufstel l ung und A nschlus s /05 0-0020 -00 20 Sic her unge n baus eits @ 2 5\m od_ 1404983 741 015 _68. docx @ 18 355 1 @ 3 @ 1
5.2.2 Customer-provided fuses
Typically, the freeze-dryer must be protected with 16 Amp G fuses that are
to be provided by the customer.