RS-232 306 Datalogger
CONTENTS TITLE I. PAGE Introduction .............................................................................................................. 1 II. Specifications ......................................................................................................... 1 III. Symbol Definition and Button Location ............................................. 2 IV. Operating Instructions .................................................................................... 3 4.1 Power-Up ..............
MARTEL 306 I. Introduction: This instrument is a digital thermometer for use with any K-type thermocouple as temperature sensor. Temperature indication follows National Bureau of Standards and IEC584 temperature/voltage table for K-type thermocouples. Its internal memory can store up to 16,312 records.(See Note 1 below.) An RS232 interface provides bi-directional communication with a PC. II.
MARTEL 306 III. Symbol Definition and Button Location: : This indicates that a negative temperature is sensed. ƨƩ K : Centigrade and Fahrenheit indication. : Thermocouple Type Indication MAX : The Maximum value is being displayed MIN : The Minimum value is being displayed : This indicates auto power off is enabled. : This indicates that the display data is being held.
MARTEL 306 IV. Operating Instructions: 4.1 Power-Up Press the power button to turn the thermometer ON or OFF. When powered on, the LCD will show how much memory space is available to use. For example: It indicates that there are 16,000 records available in memory. 4.2 Connecting the Thermocouples Plug the thermocouple into the input connectors. 4.3 Selecting the Temperature Scale When the meter is first powered on, the default scale setting is set at Celsius (°C) scale.
MARTEL 306 4.7 Recording Interval Setup : 1. Press and hold “MAX MIN” button and then power on the meter: 2. Press “HOLD" (INTV) 3. Press "REC" or "°C/°F" to increase or decrease the number. Press the “HOLD" (INTV) button to adjust next item, then press the “HOLD” (INTV) to save your changes. If you want to abort the setup process, press the power button to cancel. 4.8 Time Operation: Pressing the “TIME” button will display time.
MARTEL 306 4.13 Digital Output: The Digital Output is a 9600bps N 81 serial interface. The RX is a 5V normal high input port. The TX is a 5V normal high output port. TX RX GND The command of Digital Output is list below: RS232 command K(ASC 4BH) A(ASC 41H) H(ASC 48H) M(ASC 4DH) N(ASC 4EH) T(ASC 52H) C(ASC 43H) U(ASC 55H) P(ASC 50H) Function Ask for model No.
MARTEL 306 bit3: 1→Indicates the LCD is displaying time. bit4: not used bit5: 1→ HOLD, 0→not HOLD bit6: 1→LOW BATTERY , 0→BATTERY NORMAL bit7: 1→°C 0→°F 3th BYTE: bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0 Auto Power Off memory full resolution sign OL resolution sign OL bit0: 1→T1 is OL, 0→not OL bit1: 1→T1 value is minus, 0→T1 value is plus. bit2: 1→4th byte and 5th byte represent #### , 0→4th byte and 5th byte represent ###.# bit3: 1→T2 is OL, 0→not OL bit4: 1→T2 value is minus, 0→T2 value is plus.
MARTEL 306 V. Setup ThermoLog (Thermo DataLogger)ȋRS232 interface software: The ThermoLog package contains: 1. Two 3.5” diskettes 2. Custom designed RS232 cable for THERMOLOG. System Required: Windows 95, Windows 98, Windows NT 4.0, or above. Minimum Hardware Required: - 486-100 MHz PC-compatible, 16 MB RAM - At least 5 MB of hard disk space available to install THERMOLOG program - Display resolution of 800X600 recommended Install ThermoLog: 1.
MARTEL 306 DataLogger: Open the DataLogger Window to load the recorded data from thermometer to the PC. Tabular: Open the Tabular window to list the data from the thermometer in a scrolling table. The data can be stored as a file, or the table can be copied to other programs such as Microsoft EXCEL for further analysis. Graph: Open the Real-Time Graph window to plot the data in graph form. Exit: Terminates THERMOLOG program.
MARTEL ELECTRONICS PO Box 770, Londonderry, NH 03053 USA Tel: 800-821-0023 E-Mail: sales@martelcorp.com Web: http : / / www.martelcorp.