Table 17: RTD Read and Source
RTD Type Range (°C) Accuracy
Ni120 (672) -80.0 - 260.0 0.2
Cu10 -100.0 - 260.0 1.4
Cu50 -180.0 - 200.0 0.4
Cu100 -180.0 - 200.0 0.3
YSI400 15.00 - 50.00 0.1
Pt100 (385) -200.0 - 100.0 0.2
100.0 - 300.0 0.3
300.0 - 600.0 0.4
600.0 - 800.0 0.5
Pt200 (385) -200.0 - 100.0 0.8
100.0 - 300.0 0.9
300.0 - 630.0 1.0
Pt500 (385) -200.0 - 100.0 0.4
100.0 - 300.0 0.5
300.0 - 630.0 0.6
Pt1000 (385) -200.0 - 100.0 0.2
100.0 - 300.0 0.3
300.0 - 630.0 0.4
Pt385-10 -200.0 - 100.0 1.4
100.0 - 300.0 1.6
300.0 - 600.0 1.8
600.0 - 800.0 2.0
Pt385-50 -200.0 - 100.0 0.4
100.0 - 300.0 0.5
300.0 - 600.0 0.6
600.0 - 800.0 0.7
Pt100 (3926) -200.0 - 100.0 0.2
100.0 - 300.0 0.3
300.0 - 630.0 0.4
Pt100 (3916) -200.0 - 100.0 0.2
100.0 - 300.0 0.3
300.0 - 630.0 0.4
Read Accuracy is based on 4-wire input. For 3-wire input add ± 0.05Ω
assuming all three RTD leads are matched.
TC type Range (°C) Accuracy
U -200.0 - 0.0 0.5
0.0 - 600.0 0.25
N -200.0 - 0.0 0.8
0.0 - 1300.0 0.4
All TC errors include CJC errors
CJC error outside of 23 ± 5°C is 0.05°C/°C
(In °C add .2 for cold junction compensation error.)