Instruction Manual

99
Table 13: Resistance Measurement
Range Accuracy
(% of reading ± floor)
Ohms low 0.00W - 400.0W 0.015% ± 0.03W
Ohms high 401.0W - 4000.0W 0.015% ± 0.3W
Table 14: Resistance Source
Range Excitation Current Accuracy
(% of reading ± floor)
Ohms low 5.0W - 400.0W 0.1mA - 0.5mA 0.015% ± 0.1W
5.0W - 400.0W 0.5mA - 3mA 0.015% ± 0.03W
Ohms high 400W - 1500W 0.05mA - 0.8mA 0.015% ± 0.3W
1500W - 4000W 0.05mA - 0.4mA 0.015% ± 0.3W
Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <=
5ms.
Table 15: Thermocouple Measurement/Source
Range Accuracy
(% of reading ± floor)
Read (mV) -10.000mV - 75.000mV 0.02% ± 10mV
Source (mV) -10.000mV - 75.000mV 0.02% ± 10mV
Maximum current output in voltage ranges is 1mA with an output impedance of <= 1W
Table 16: Thermocouple Read and Source (errors in °C)
TC Type Range (°C) Accuracy
J -210.0 - 0.0 0.4
0.0 - 800.0 0.2
800.0 - 1200.0 0.3
K -200.0 - 0.0 0.6
0.0 - 1000.0 0.3
1000.0 - 1372.0 0.5
T -250.0 - 0.0 0.6
0.0 - 400.0 0.2
E -250.0 - -100.0 0.6
-100.0 - 1000.0 0.2
R 0.0 - 1767.0 1.2
S 0.0 - 1767.0 1.2
B 600.0 - 800.0 1.2
800.0 - 1000.0 1.3
1000.0 - 1820.0 1.5
C 0.0 - 1000.0 0.6
1000.0 - 2316.0 2.3