Owner manual
Output Specifications
Function
Range
Resolution
Accuracy
B
Volts
0-1 Vdc
10 µV
±0.005%
RDG ±0.005
A
0-10 Vdc
100 µV
±0.005% RDG ±0.005
A
Impedance:
<5 ohms
mV
0-100 mVdc
1 µV
± 0.005% RDG ±0.005
A
Impedance:
<5 ohms
mA
0-22 mA
1 µA
±0.005% RDG ±0.01
A
Impedance:
>1 megohm
Freq
0.01-9.9999 Hz
0.0001 Hz
C
±0.005% RDG
D
10-99.999 Hz
0.001 Hz
C
±0.005% RDG
D
100-999.99 Hz
0.01 Hz
C
±0.005% RDG
D
1000-9999.9 Hz
0.1 Hz
C
±0.005% RDG
D
10000-100000 Hz
1 Hz
C
±0.005% RDG
D
Amplitude:
250mV to 10V Peak
Rise and Fall:
1µsec (10%-90%)
Waveform:
Ground based positive square wave
Lock Time:
1 second to within 1% of final frequency
Impedance:
100 ohms nominal
A
% of full scale +/- 1 digit.
B
Accuracy stated at 25°C, traceable to NIST.
C
Fully Autoranging.
D
When measured with a Gate Time of 1 second.