Owner manual
Specifications
Input Specifications
Function
Range
Resolution
Accuracy
B
Volts
0-1 Vdc
10 µV
±0.005%
RDG ±0.005
A
0-10 Vdc
100 µV
±0.005% RDG ±0.005
A
Impedance:
>10 megohms nominal
Max. Input:
30 Vdc
mV
0-100 mVdc
1 µV
± 0.005% RDG ±0.005
A
Impedance:
>10 megohms nominal
mA
0-20 mA
1 µA
±0.005% RDG ±0.01
A
0-50 mA
10 µA
±0.005% RDG ±0.01
A
Impedance:
5 megohms nominal
Freq
5-9.9999 Hz
0.0001 Hz
C
±0.005% RDG
10-99.999 Hz
0.001 Hz
C
±0.005% RDG
100-999.99 Hz
0.01 Hz
C
±0.005% RDG
1000-9999.9 Hz
0.1 Hz
C
±0.005% RDG
10000-100000 Hz
1 Hz
C
±0.005% RDG
Amplitude:
50mV to 170V Peak
Sensitivity:
50mV RMS 5Hz to 20kHz
100mV RMS 20kHz to 100kHz
Reading Rate:
2 readings per second
Impedance:
1 megohm nominal
A
% of full scale +/- 1 digit.
B
Accuracy stated at 25°C, traceable to NIST.
C
Fully autoranging.
Specifications 130/135 67