Owner`s manual
14
SECTION 5 FIELD SERVICE INSTRUCTIONS & ASSEMBLY
Door Figure 1.
STERILMATIC OPEN STAND:
Market Forge Sterilmatic Stand can
be supplemented with an Optional
Stand for utility use where maxi-
mum compactness is desired.
The sturdy, stainless steel unit is
equipped with adjustable leg ex-
tensions which allow the unit to be
installed and leveled over existing
contours in the oor.
The open design lends itself to maximum sanitary con-
ditions because of the ease with which periodic clean-
ing can be done.
Though simple in design and appearance, the steril-
matic stand is the ideal arrangement for mounting in
that it allows secondary air to circulate.
STERILMATIC OPEN STAND WITH
CONDENSER:
Market Forge can provide the open stand with an op-
tional steam condenser system for use where steam
exhaustion into the room is undesirable.
The condenser is automatically controlled by the ther-
mostat. The normal factory thermostat setting is 130°F
(54°C). The open under-shelf of the stand gives added
utility providing a handy tabouret for utensils and ac-
cess for drainage of water from the sterilizing cham-
ber.
PARTS LIST FOR CONDENSER WITH
OPTIONAL STAND:
PART NO. 50 Hz DESCRIPTION
10-4653 10-4653 Thermostat
10-4035 10-7074 3/8” Solenoid
10-5731 10-5731 1/2” Water Stop Valve
95-2106 95-2106 Water Injection Assy.
95-1680 95-1680 Shelf
STERILMATIC DOOR ASSEMBLY:
The Door of the Sterilmatic has been engineered to
establish a positive method of sealing the steam pres-
sure within the sterilizing cylinder. As steam pressure
builds up within the cylinder, the door seal will tend to
become more positive.
However, the door should be adjusted to make a good
initial seal between the door gasket and the door open-
ing without the added assistance of internal cylinder
steam pressure with the simple action of securing the
door handle down in a locked position, the door gas-
ket should be sufciently compressed against the door
opening, all the way around to prevent any steam leak-
age from occurring.
DOOR ADJUSTMENT:
The Door Adjustment is Located in the Fulcrum Cast-
ing at the base of the door opening. This adjustment
employs the use of a screw and locknut in order to ad-
just the Sterilmatic Door to a tighter closed position (to
prevent steam from leaking by the door gasket as pres-
sure builds up), it is necessary to loosen the locknut
and back off the screw at least one-quarter of a turn
and re-tighten the locknut (see Door gure 1 shown
above & Door Figure 2 shown below).
Door Figure 2.