Owner`s manual

12
do not recommend the sterilization of large packs,
such as linens. Be sure condensate bafes are in
position in the chamber.
Remove load and check water level for next opera-
tion.
STERILIZATION GUIDE:
PACKS (Linens, gloves, etc.): Use wire basket
to facilitate drying. Be sure condensate bafes are
in place. Place packs on edge and arrange load in
chamber, so that only minimal resistant to passage
of steam through the load will exist. NOTE: Place
gloves in upper two-thirds of chamber.
JARS, CANISTERS (etc.): Place containers on
side to allow for displacement of air and complete
contact of steam to surfaces. Drying is also facili-
tated.
PETRI DISHES, PIPETTES, DESICCATORS
(etc.): Should be inverted.
UTENSILS, TREATMENT TRAYS: Placed on
edges to facilitate drying.
INSTRUMENT SETS: Place instruments set in
9.
trays having mesh or perforated bottoms. Place
trays at on shelves.
COMBINING FABRICS & HARD GOODS:
Place hard goods on lowest shelves.
PLASTIC UTENSILS: DO NOT stack or nest
plastic items.
LIQUIDS: Sterilize medium liquids separately
from other supplies or materials. Set exhaust se-
lector to proper position (liquids).
SMALL ITEMS: Sterilize small items in baskets,
or trays.
SECTION 3 OPERATING INSTRUCTIONS
MINIMUM STERILIZATION TIMES
TIME (Minutes) ARTICLES
15 Glassware, empty, inverted.
Instruments, metal in covered or open tray, padded or unpadded.
Needles, unwrapped.
Pipettes, blood diluting, serological, volumetric, etc
Tubing glass (6mm), (10mm) inverted
20 Flasked solutions 75-250 ml.
Instruments, metal combined with other materials in covered and/or padded tray.
Instruments wrapped in double thickness muslin.
Rubber gloves, catheters, drains, tubing, etc. Unwrapped or wrapped in muslin or
paper.
30 Brushes in dispensers, in cans of individually wrapped.
Dressings, wrapped in paper or muslin, small packs only.
Flasked solutions 500-1000 ml.
Syringes, unassembled, individually packaged in muslin or paper.
Needles, luer, individually packaged in glass tubes or paper.
45 Flasked solutions 1500-2000 ml.
NOTE: IF THE EQUIPMENT IS USED IN
A MANNER NOT SPECIFIED BY THE
MANUFACTURER, THE PROTECTION
PROVIDED BY THE EQUIPMENT MAY BE
IMPAIRED.