SECTION 1 INSTALLATION INSTRUCTIONS
INSTALLATION FOR OPTIONAL RECORDING
THERMOMETER: The optional Recording Thermom-
eter may be surface or panel mounted whichever is
preferred.
Remove the top cover of the sterilizer. Remove
the pipe plug located in the T-tting that also ac-
commodates the safety relief valve and which is
connected directly to the sterilizing chamber. (See
gure to the right)
Insert the RTD temperature sensor of the tempera-
ture recorder through the rectangular cutout in the
top cover then into the pipe from which the plug
was removed.
Make sure RTD temperature sensor is protruding
down into the sterilizer cavity but not too much so
as to interfere with the sheet metal bafe plate.
Tighten the compression tting. Replace top cover
onto the sterilizer.
For mounting the recorder on a wall, refer to the
recorder manual supplied with the recorder.
Electrical connections: The 115/120VAC unit comes
with a 6’ power cord and three prong plug. This can
be plugged into any 120V outlet. The 220/230VAC
unit comes with a 6’ power cord and three prong
plug. However, due to the many varieties of plugs
and outlets you may need to replace the plug with
the appropriate plug.
The Future Design Recording Thermometer is a 24-
hour clock and should be run continuously.
Charts should be changed every 24 hours.
1.
2.
3.
4.
5.
TO REMOVE THE CHART:
Unscrew and remove the chart ‘hub’ knob (at-
tached to the chain).
Grasp the edge of the chart and lift straight off.
TO INSTALL A NEW CHART:
Slide edge of the chart under the two raised tabs
located under the stylus.
Locate the center hole of the chart over the cen-
ter, threaded rod. Rotate chart to line up the cur-
rent time of day directly underneath the stylus print
head.
Screw the chart ‘hub’ knob back in place, hand
tight only.
1.
2.
1.
2.
3.