STERILIZER TROUBLE SHOOTING GUIDE
SYMPTOM POSSIBLE CAUSE CORRECTION P/N
Sterilmatic reaches correct 1) Steam trap not venting 1) Clean or replace thermo-element
10-1092 ( see page 18 )
pressure but not correct properly in trap
temperature 2) Temperature gauge not 2) Replace temperature gauge
10-9268
reading properly
Sterilmatic continues to 1) Pressure control contacts 1) Replace defective switch(es)
95-3442 (STM-E) ( see page 12 )
maintain pressure when timer remain closed
is in off position
95-3434(STM-EL) ( see page 14 )
Sterilized packs contain 1) Condensate or water baffle 1) Verify both interior baffles are in place, if
excessive moisture missing not then install as per (pg. 6) inst. instr.
2) Steam trap is closing 2) Clean or replace thermo-element in
10-1092 ( see page 18 )
prematurely, trapping air in steam trap
chamber
* INDICATES DIFFERENT P/N FOR EXPORT UNITS