Operating instructions
19
SECTION 5 ASSEMBLY
TO REMOVE THE DOOR ASSEMBLY
The Door Assembly can be removed from the inner sterilizing chamber as a unit without the
use of any special tools or equipment. However, a systematic approach to this is warranted
as the clearances through the portal are close, and much confusion can result if not removed
in the sequence described below:
1. First, lift off and remove the two pan supports to expose the door linkage on either side of the
inner sterilizing chamber .
2. Raise the door to a fully opened position, and disengage the door spring from each of the door
spring studs. Accomplish this by counteracting the force of the door lift spring with one hand while
working the end of the door spring off the spring stud with the free hand. Do this on both sides of
the door assembly.
3. When the end of the door springs have been completely freed from their respective door spring
studs, the door springs on either side of the door assembly can easily be slipped off their studs.
4. Rotate the entire door assembly out through the door opening, passing the door handle through
the opening first, and then one end of the door spring as shown in the illustration. The remainder
of the door assembly will then pass through the door opening quite easily.
5. To replace the door assembly, reverse the step-by-step procedure described above.
ITEM PART NO. DESCRIPTION
1 10-6765 Pivot Spring Bearing
2 91-2718 Right and Left Door Spring (sold as a pair)
3 10-1776 10-32 Machine Screw 1/2" Long
4 10-2666 Door Gasket
5 95-3204 Door & Door Spring Assembly
95-0124 ITEMS 1 THROUGH 6