To save the calibration information, select “Save & exit”. To exit without saving the data select
“Exit without saving”.
12. Any errors are reported by the following screens:
Displayed at the start of calibration if a disallowed unit is selected.
Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.
Causes:
1. The calibration weight does not match the set value.
2. If using a PTA configurable adapter, ensure that the output signal leads (SG+ and SG-) have
been installed into the appropriate terminal blocks. Some sensor manufacturers consider
SG+ to be a compression value, while others consider it to be a tension value. If the indicator
is expecting a compression load but it is receiving a tension signal, calibration cannot
continue. Verify that the tension/compression indicator on the home screen properly
corresponds to the load direction, and switch the signal leads, if required.
CALIBRATION
Load not stable.
Please try again.
CALIBRATION
Units must be gF.
Please try again
Press ENTER.
CALIBRATION
COMPRESSION
Load too low.
Please try again.
CALIBRATION
COMPLETE
Save & exit
Exit w/o saving