Specifications

M-Series Water Purification Machines
1238160 Rev. D 1 - 8 Description
Autoflush System – programmable, automated high velocity membrane
flushing for the longest membrane life; set at the factory and adjustable in
the field, includes a digital panel-mount timer, complete with push button
Manual Flush
An electrical package that includes alarm delay shutdown for low inlet
pressure condition to prevent pump damage should pressure fall below 15
psig (1 bar)
Conductivity monitor, panel-mounted, for permeate quality monitoring
NEMA-4X fiberglass electrical enclosure
1.5 Specifications for M-Series Machines
1.5.1 Feedwater
TEMPERATURE:
35-77° (22-25°C) [Not to exceed 85°F (29°C) unless specifically designed for
higher temperatures]
INLET PRESSURE:
Minimum: 30 psig (2.1 bar)
Maximum: 60psig (4.1bar)
CHLORINE (CONTINUOUS FEED):
For FASTEK TLC
membranes 0 ppm
OPERATING pH:
Softwater [less than 1 grain per gallon 3.0-10.0 (gpg) or 17 mg/L hardness]
Unsoftened water (contact factory 5.5-6.0 with water analysis)
PRE-FILTER:
5-micron HYTREX cartridge (see machine label for part number)
INLET CONNECTIONS:
¾-inch FNPT (IPS)