Specifications
  23G-Series RO System 
3024270 Rev G  83  14Oct13 
Table 6.10 Two-Digit HyperTerminal Alarm Codes 
Number on Left 
Number on Right 
0 No Alarm 
0 No Alarm 
1 Metric 
1 Permeate Conductivity High 
2 Permeate Divert 
2 High Temperature 
3 Metric and Divert 
3 High pH 
4 HAND mode 
4 Low pH 
5 Metric and HAND mode 
5 Low Post Filter Pressure 
6 Permeate DIVERT and HAND mode 
6 High Primary Pressure 
7 Metric and Divert and HAND mode 
7 Low Primary Pressure 
8 CIP 
8 High Permeate Pressure 
9 Rejection Low 
A Prefilter Delta P 
B Membrane Delta P 
C Permeate Flow Low 
D Tank Full 
E Pretreatment Lockout 