8.  Rinse or blot away any remaining droplets of the pH solution and 
reinstall the pH sensor into the incoming waterline. 
9.  Generally this 2-point pH calibration procedure should be repeated 
on a bi-monthly basis. Some situations will require more frequent 
recalibration while others will require less. 
10.  If the probe does not seem to be responding, try soaking it in a 2.0 
pH acid for 10 minutes. Repeat the calibration. 
11.  If the probe will not calibrate after 2 or 3 attempts, refer to the pH 
section in the Troubleshooting Guide (Section 7.0). 
12.  Save the settings using passcode 2 in the maintenance menu 
Factory Settings. 
8.3  V2 Logic Board 
NOTE:   
For machines equipped with Rev G2.2 software as shipped (V2 Logic 
Board). These documented specifications are subject to change, 
depending on the unique characteristics of each machine during quality 
assurance (QA) testing. Any “user” variation of these specifications 
(including any software upgrades) in the field (after the machine has left 
Mar Cor Purification shipping dock) is solely the responsibility of the user 
to document and control. 
Menu #2 
Alarms 
ECM Adjustment Range 
Alarm Limit 
PERM COND HI 
0-200 S/cm
ab 
30 S/cm 
TEMP HI
c
0-255°F (0-123°C) 
90°F (32°C) 
pH HI
c 
0.0-14 pH units
d 
9.0 pH units 
pH LOW
c 
0.0-14 pH units 
3.5 pH units 
POST-FILTER PRESS LOW
c 
0-99 psi (0-68 [barX10]) 
12 psig (8 [barX10]) 
PRIMARY PRESS HI
e 
0-510 psi (0-352 [barX10]) 
280 psig (193 [barX10]) 
PRIMARY PRESS LOW
c 
0-510 psi (90-352 [barX10]) 
150 psig (103 [barX10]) 
PERM PRESS HI
e 
0-99 psi (0-68 [barX10]) 
80psig (55 [barX10]) 
REJECTION LOW 
0-99% 
88% 
FILTER PRESS DROP HI 
0-99 psi (0-68 [barX10])
f 
14 
MEMBRN PRESS DROP HI 
0-510 psi (0-352 [barX10])
g 
108 
PERM FLOW LOW 
0.0-20.1 gpm (0.0-6.5L)
h 
2.6 