Specifications
  23G-Series RO System 
3024270 Rev G  101  14Oct13 
The ECM-100 calculates % rejection using the following equation: 
% Rejection =  
1-  
Permeate Conductivity 
 x 100 
Feed Conductivity 
Since the % rejection is based on a ratio between the feed 
conductivity and the permeate conductivity, higher quality feed 
water may result in a low % rejection alarm. Higher recovery will 
typically lower the permeate rejection using the ECM- 100 
calculations. 
If a low % rejection alarm results from high purity feed water when 
the permeate conductivity is at acceptable levels, the following 
three methods could be used: 
1.  Reduce low % rejection alarm setpoint. 
2.  Adjust the feed conductivity calibration potentiometer for a 
slightly higher feed conductivity reading. See Section 8.0 for 
calibration of feed conductivity. 
NOTE:   
Adjusting feed conductivity to a slightly higher reading will result in an 
inaccurate feed conductivity displayed on the ECM-100. 
3.  Use a resistor in place of the feed sensor. Remove the feed 
conductivity input and replace with a resistor to simulate a feed 
conductivity. A 10 K ohm resistor would indicate 250 S. 
NOTE:   
To properly monitor permeate conductivity make sure the PERM HI COND 
ALARM Menu in the Maintenance Menu (10) is properly set. See Section 
4.4.10 for accessing Permeate High Conductivity Alarm Menu. 
For additional assistance contact your equipment supplier. 










