Service manual
Checking the universal stand on/off switch connections
u Shown to the right, is the On/Off switch installed in the Universal Stand.
u Using a small flat blade screwdriver, gently pry the switch out.
u Once the switch is loose, pull it out using your fingers.
u Verify that both wires are properly connected (the switch shown has one wire
disconnected).
u With both wires connected, reinstall the switch by gently pressing
it back into place.
u The switch should look as shown when properly reinstalled.
www.visioneng.com/support Mantis Service Manual