Service manual

Section 6: Disassembly Guide
6-7
6. Position the ground line from the PEM so that it does not come into contact
with components on the Power Supply PCB.
7. Complete procedure in paragraph 6.5.
6.8 POWER SUPPLY REMOVAL/INSTALLATION
Removal
1. Complete the procedure described in paragraphs 6.2 and 6.4.
2. Disconnect the leads from the battery.
3. Follow the procedure in paragraph 6.7, steps 2 and 3.
4. Use a 10mm wrench to disconnect the Power Supply ground lead from the
equipotential lug (Figure 6-6).
5. Remove the seven screws shown in Figure 6-7.
6. Lift the Power Supply out of the bottom case.
W1
to
Equipotential
Lug
W2
Brown to
"L" on
PEM
W3
Blue to
"N" on
PEM
W5
Black
W4
Red
Equipotential
Lug
G
LN