Service manual
4-1
SECTION 4: AUDIBLE ALARM SETTINGS & SERVICE MENU
4.1 Introduction
4.2 Audible Alarm Settings
4.3 Service Menu
4.1 INTRODUCTION
This section discusses use of the service menu to reconfigure power-on default
values, and how to control the behavior of the audible alarm.
4.2 AUDIBLE ALARM SETTINGS
The following paragraphs describe how to change the behavior of the audible
alarm. Operators can select the volume of the alarm and the duration of alarm
silence. Controls for the NPB-190 are shown in Figure 4-1.
NPB190
On/Standby Alarm silence
Adjust
up
Adjust
down
Set lower
limit
Set upper
limit
Figure 4-1: NPB-190 Controls
4.2.1 Alarm Silence State
Press the Alarm Silence button to silence the alarm. Press the button a second
time to turn the alarm back on.
4.2.2 Alarm Silence Duration
1. Press and hold the Alarm Silence button for less than 3 seconds.
2. Before 3 seconds have passed the Adjust Up or Adjust Down button can be
used to change the duration of the alarm silence. The alarm’s duration can
be 30, 60, 90, 120 seconds, or the alarm can be turned off.
4.2.3 Alarm Volume
1. Press and hold the Alarm Silence button for more than 3 seconds.