OPERATOR’S MANUAL NPB-290 Pulse Oximeter Caution: Federal law (U.S.) restricts this device to sale by or on the order of a physician. To contact Mallinckrodt’s representative: In the United States, call 1.800.635.5267 or 314.654.2000; outside of the United States, call your local Mallinckrodt representative. 2002 Mallinckrodt Inc. All rights reserved.
Nellcor Puritan Bennett Inc. 4280 Hacienda Drive Pleasanton, CA 94588 USA Telephone Toll Free 1.800.NELLCOR Tyco Healthcare UK LTD Fareham Road Gosport PO13 0AS U.K. Tel: +44.1329.224000 Nellcor Puritan Bennett is a wholly owned subsidiary of Mallinckrodt Inc. Nellcor and Nellcor Puritan Bennett are trademarks of Mallinckrodt Inc. To obtain information about a warranty, if any, for this product, contact Mallinckrodt’s Technical Services Department, or your local Mallinckrodt representative.
CONTENTS Figures Tables CONTENTS CONTENTS ............................................................................... Contents .......................................................................... Figures............................................................................. Tables.............................................................................. SAFETY INFORMATION .......................................................... General Safety Information ..............................
Contents List of Components ......................................................... Optional Accessories............................................ Monitor Setup.................................................................. General................................................................. Setting Up the NPB-290 ....................................... SENSORS AND ACCESSORIES............................................. Selecting a Sensor..........................................................
Contents Returning The NPB-290 ....................................... Maintenance .................................................................... Service .................................................................. Periodic Safety Checks......................................... Performance Verification ...................................... Cleaning ................................................................ SPECIFICATIONS.....................................................................
Contents Trend Data Operation........................................... 69 Nurse Call ....................................................................... 70 Analog Outputs ...............................................................
Contents FIGURES Figure 1: Figure 2: Figure 3: Figure 4: NPB-290 Front Panel Display .................................. NPB-290 Rear Panel ................................................ Connecting the Power Cord ..................................... Connecting an MC-10 or SCP-10 Sensor Cable to the NPB-290............................................... Figure 5: Attaching the Mounting Plate ................................... Figure 6: Attaching the Tilt-Stand Supports.............................
(Blank Page)
SAFETY INFORMATION General Safety Information GENERAL SAFETY INFORMATION This section contains important safety information related to general use of the NPB-290 pulse oximeter. Other important safety information appears throughout the manual in sections that relate specifically to the precautionary information. Be sure to read all text surrounding all precautionary information.
Safety Information compliance with IEC Standard 601-1-1 systems requirements. Anyone who connects additional equipment to the signal input port or signal output port configures a medical system and is therefore responsible that the system complies with the requirements of system standard IEC Standard 601-1-1 and the electromagnetic compatibility standard IEC Standard 601-1-2. To ensure accurate readings, consider the environmental conditions that are present and the condition of the patient.
INTRODUCTION Intended Use General Operating Principles and Conditions INTENDED USE The Nellcor NPB-290 pulse oximeter is intended for continuous, noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), and pulse rate (measured by SpO2 sensor). The monitor is intended for use on adult, pediatric, and neonatal patients in all hospital areas, hospital-type facilities, and in the home environment. It may be used during intra-hospital transport when powered by its internal battery.
Introduction Specific information about ambient environmental conditions, sensor application, and patient conditions, is contained throughout this manual.
CONTROLS, INDICATORS, AND SYMBOLS Displays, Controls, Indicators, and Connectors Other NPB-290 Symbols Description of Controls Description of Visible Indicators and Displays Description of Audible Indicators DISPLAYS, CONTROLS, INDICATORS, AND CONNECTORS Figures 1 and 2 show the front and rear views of the NPB-290 and identify displays, controls, and connectors.
Controls, Indicators, and Symbols 3 2 1 NPB-290 SN 2X NELLCOR PURITAN BENNETT, INC. PLEASANTON, CA 94588, U.S.A. T 0.50A 250V NELLCOR PURITAN BENNETT EUROPE BV, 's-HERTOGENBOSCH, THE NETHERLANDS R IPX1 012 3 NRTL/C 5 1 2 3 4 5 100-120 V 200-240 V 50/60 Hz 20VA MADE IN U.S.A. 035285X-1296 U.S.
Controls, Indicators, and Symbols DESCRIPTION OF CONTROLS Function Buttons The Power On/Off button. Used to turn the NPB-290 monitor on or off. The Alarm Silence button. Used to silence current alarms for the alarm silence duration period. It is also used to view and adjust alarm silence duration and the alarm volume. The Adjust Up button. Used to increase alarm limit values, alarm silence duration, as well as alarm and pulse beep volumes. Also used to scroll through menu items. The Adjust Down button.
Controls, Indicators, and Symbols The Pulse Search indicator. Lights continuously prior to initial acquisition of a pulse signal, and during Pulse Search mode. It flashes during loss of pulse. DESCRIPTION OF VISUAL INDICATORS AND DISPLAYS The %SpO2 Display. Shows the hemoglobin oxygen saturation level. The Pulse Amplitude indicator. A 10-segment display that shows the relative pulse amplitude. As the detected pulse becomes stronger, more contiguous segments light with each pulse.
Controls, Indicators, and Symbols Pulse Beep A single beep sounds for each detected pulse. Confirmation Tone Three beeps sound to indicate a successful menu operation. Alarm Silence Reminder Three beeps sound at least every 3 minutes as a reminder that the audible alarm has been silenced. Volume Setting Tone A continuous tone that is used to adjust the alarm volume.
(Blank Page)
QUICK GUIDE TO OPERATION Basic Operation Alarm Functions This Quick Guide to Operation section is intended for use by experienced NPB-290 users. First-time users of this instrument should read the entire manual before use. BASIC OPERATION The NPB-290 operates on AC power when the hospital-grade power cord is connected to both the monitor and an AC power source (wall outlet). The monitor will not operate when its internal battery is discharged, even when the monitor is connected to AC power.
Quick Guide to Operation 3. Select a Nellcor sensor that is appropriate for the patient being monitored. 4. Attach the sensor to the patient, following the instructions in the sensor directions for use. WARNING: Do not use an extension cable with the NPB-290 monitor (other than the MC-10 or SCP-10 sensor cable). Use of another sensor extension cable will have an adverse effect on performance. Do not attach to the sensor port any cable that is intended for computer use. 5.
Quick Guide to Operation 3. When a valid pulse is detected, the NPB-290 begins monitoring. ADJUST PULSE BEEP VOLUME The pulse beep volume can be adjusted by pressing either the Adjust Up or Adjust Down buttons located on the right side of the front panel. ALARM FUNCTIONS View/Adjust Alarm Limits To view the current upper alarm limit for SpO2, press the upper alarm limit button once. The upper alarm limit is shown in the %SpO2 Display.
Quick Guide to Operation Adjust Alarm Silence Duration To adjust the alarm silence duration, press and hold the Alarm Silence button. The current alarm silence duration is shown in the display. Press the Adjust Up or Adjust Down button to change the silence duration to the desired amount of time. Disabling Audible Alarms Setting the alarm silence duration to OFF means that no audible alarms will be produced by the monitor.
SETUP Unpacking and Inspection Performance Verification List of Components Monitor Setup UNPACKING AND INSPECTION Notify the carrier if the shipping carton is damaged. Unpack the NPB-290 and components. If anything is missing or damaged, contact Mallinckrodt’s Technical Services Department or your local Mallinckrodt representative.
Setup MONITOR SETUP General WARNING: To ensure patient safety, do not place the monitor in any position that might cause it to fall on the patient. WARNING: As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation. WARNING: Disconnect the NPB-290 and Nellcor sensor from the patient during magnetic resonance imaging (MRI) scanning. Induced current could potentially cause burns.
Setup The supply voltage selector switch allows connection of the monitor to AC power ranging from 100 VAC to 240 VAC. The switch has two positions: one for 100-120 VAC (“115”), and one for 200-240 VAC (“230”). Ensure that the supply voltage selector switch on the rear panel is set to the proper voltage. Operating on a discharged battery The NPB-290 will not operate when its internal battery is discharged, even when the monitor is connected to AC power. Instead, the error code “EEE 04” is displayed.
Setup Figure 3: Connecting the Power Cord 3. Verify that the supply voltage selection switch located on the rear of the monitor matches the AC power requirement at your location. 4. Plug the male connector end of the power cord into a properly grounded AC outlet. WARNING: In the USA, do not connect the monitor to an electrical outlet controlled by a wall switch. 5.
Setup 7. Plug the MC-10 or SCP-10 cable into the sensor port located on the front of the NPB-290 (as shown in Figure 4).
(Blank Page)
SENSORS AND ACCESSORIES Selecting a Sensor Biocompatibility Testing Performance Considerations Optional Accessories SELECTING A SENSOR WARNING: Before use, carefully read the sensor directions for use, including all warnings, cautions, and instructions. WARNING: Do not use a damaged sensor or sensor cable. Do not use a sensor with exposed optical components. WARNING: Use only Nellcor sensors and sensor cables with this monitor. Other sensors or sensor cables may cause improper NPB-290 performance.
Sensors and Accessories Table 1: Nellcor Sensors Sensor Model Patient Size Oxisensor® II oxygen transducers (Sterile, single-use only) N-25/N-25LF I-20/I-20LF D-20 D-25(L) R-15 <3 or >40 kg 3 to 20 kg 10 to 50 kg >30 kg >50 kg Oxiband® oxygen transducer (Reusable with disposable nonsterile adhesive) OXI-A/N OXI-P/I <3 or >40 kg 3 to 40 kg Durasensor® oxygen transducer (Reusable, nonsterile) DS-100A >40 kg Nellcor reflectance oxygen transducer (reusable, nonsterile) RS-10 >40 kg Dura-Y ® mult
Sensors and Accessories Inaccurate measurements can be caused by: • incorrect application of the sensor • placement of the sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line • ambient light • patient movement • Loss of pulse signal can occur for the following reasons: • the sensor is too tight • a blood pressure cuff is inflated on the same extremity as the one with the sensor attached • there is arterial occlusion proximal to the sensor Use only Nel
Sensors and Accessories • Use a new sensor with fresh adhesive backing. If poor perfusion affects performance, consider using the Oxisensor R-15 sensor. It obtains measurements from the nasal septal anterior ethmoid artery, an artery supplied by the internal carotid. This sensor may obtain measurements when peripheral perfusion is relatively poor.
Sensors and Accessories Figure 5: Attaching the Mounting Plate Tilt-Stand Supports A pair of tilt-stand supports are available from Mallinckrodt for the NPB-290 monitor. These supports provide the option of tilting the front of the monitor upward in order to view the display at an angle. When not in use, the supports can be folded flat against the bottom of the monitor. The supports are attached to the NPB-290 as shown in Figure 6.
Sensors and Accessories Figure 6: Attaching the Tilt-Stand Supports 26
START-UP AND USE Basic Operation Alarms Trend Data Battery Operation Disposal of Device Components Performance Considerations BASIC OPERATION WARNING: The NPB-290 is a prescription device and is to be operated by qualified personnel only. WARNING: Do not lift the monitor by the sensor cable or power cord because the cable or cord could disconnect from the monitor, causing the monitor to drop on the patient. WARNING: The NPB-290 is intended only as an adjunct in patient assessment.
Start-Up and Use Ensure that the voltage selector switch on the rear panel is set to the proper voltage. Power-On Self-Test (POST) WARNING: Ensure that the speaker is clear of any obstruction. Failure to do so could result in an inaudible alarm tone. 1. Plug an appropriate Nellcor sensor firmly into the MC-10 or SCP-10 and lower the MC-10 or SCP-10 sensor lock over the sensor connector until it clicks into place. Insert the MC-10 or SCP-10 into the NPB-290 sensor port.
Start-Up and Use the numbers and have them available prior to requesting technical assistance. 5. If the NPB-290 detects an internal problem during the selftest, an error tone sounds and the monitor displays an Error Code and corresponding number (see the Troubleshooting section). 6 Upon successful completion of the self-test, the NPB-290 sounds a 1-second tone indicating that the monitor has passed the test. WARNING: If you do not hear the POST pass tone, do not use the monitor.
Start-Up and Use Monitoring During monitoring, the NPB-290 displays SpO2 and pulse rate readings with each pulse beat. %SpO2 is displayed for values between 0% and 100%. Pulse rates are displayed for values from 20 to 250 beats per minute, and for zero beats per minute. Note: Pulse rates below 20 (except zero) are displayed as 20, and pulse rates above 250 are displayed as 250. A pulse rate of zero is used to indicate that the monitor is not monitoring a pulse.
Start-Up and Use At Initial Power-Up Immediately after the NPB-290 displays its software version numbers and POST is completed, the monitor enters Pulse Search (the Pulse Search indicator lights), if a sensor is properly connected to the monitor. During this Pulse Search, the display reads zeroes while the monitor searches for a valid pulse. When a valid pulse is detected, the NPB-290 begins monitoring.
Start-Up and Use ALARMS Description of Alarms The NPB-290 has three levels of audible alarms. 1. High-priority alarm: Indicated by a fast-rate, high-pitched, pulsing tone. A high-priority alarm sounds after loss-ofpulse is detected. Refer to the “Management of the Loss-of-Pulse Alarm” paragraph later in this section for a detailed description of the loss-of-pulse alarm. 2. Medium-priority alarm: Indicated by a medium-rate, medium-pitched, pulsing tone.
Start-Up and Use Adjusting Alarm Limits WARNING: Each time the monitor is used, check alarm limits to ensure that they are appropriate for the patient being monitored. Overview When the NPB-290 is first powered on, alarm limits are set to their power-on default values. Power-on default alarm limits can be changed by qualified service personnel, using the instructions described in the NPB-290 service manual. You can temporarily change alarm limits from their default values if necessary, as described below.
Start-Up and Use To view the current lower alarm limit value for SpO2, press the Lower Alarm Limit button. The number displayed in the %SpO2 field represents the current SpO2 lower alarm limit. A series of dashes are displayed in the Pulse Rate field. Since the parameter being displayed is a lower limit, these dashes appear in the lower portion of the Pulse Rate display, as shown: To view the current upper alarm limit value for Pulse Rate, press the Upper Alarm Limit button twice.
Start-Up and Use To change an existing alarm limit, first enter the viewing mode for that value, as described earlier in “Viewing Current Alarm Limits.” Once the value you wish to change is displayed, press the Adjust Up or Adjust Down button to adjust the value as needed. Alarm Limits Changed Indicator If alarm limits are changed from the NPB-290’s power-on defaults, a decimal point appears after the displayed value, during patient monitoring, or when alarm limits are viewed.
Start-Up and Use Viewing Current Alarm Silence Duration To view the current audible alarm silence duration, press and hold the Alarm Silence button for less than 3 seconds. The display will show the current duration in seconds, as shown: To adjust the audible alarm silence duration, while holding down the Alarm Silence button, use the Adjust Up or Adjust Down button to set the duration to 30, 60, 90, or 120 seconds, or OFF (Alarm Silence mode).
Start-Up and Use Alarm Silence Reminder The alarm silence reminder (three beeps) sounds at least every 3 minutes while the alarm silence duration is set to OFF. The alarm silence reminder may also be disabled by service personnel using the procedures described in the service manual. The factory default is that the reminder is enabled.
Start-Up and Use alarms are distracting and potentially time-consuming. However, if interference persists, an alarm should notify the attendant. The NPB-290 loss-of-pulse alarm is designed for use in environments in which the pulse can be obscured by interference. Most loss-of-pulse alarms are attributable to interference from patient motion, which typically is brief.
Start-Up and Use SPORADIC OR LOST PULSE WITH CONTINUOUS MOTION Low priority alarm sounds Pulse Search Indicator lights continuously Pulse Search Indicator flashes %SpO2 and Pulse Rate Displays alternate between previous display and dashes Zeros displayed Motion Indicator lights continuously* 0 seconds 10 seconds 60 seconds 50 seconds SPORADIC OR LOST PULSE WITHOUT CONTINUOUS MOTION High priority alarm sounds Pulse Search Indicator lights continuously Pulse Search Indicator flashes %SpO2 and Puls
Start-Up and Use During that data evaluation period: • If the patient is not moving (for example, because of cardiac arrest, shock, paralysis, anesthesia) and has no qualified pulse for 6 seconds, a high-priority alarm immediately sounds, the Pulse Search indicator flashes, and the data displays flash zeroes. • If the patient is constantly moving, the NPB-290 searches for qualified pulses for up to 50 seconds. Each time a qualified pulse is detected, even during motion, the data displays are updated.
Start-Up and Use Nurse Call Feature WARNING: The nurse call feature should not be used as the primary source of alarm notification. The audible and visual alarms of the monitor, used in conjunction with clinical signs and symptoms, are the primary sources for notifying medical personnel that an alarm condition exists. The nurse call feature of the NPB-290 works in conjunction with the nurse call system of your institution when the monitor sounds an audible alarm.
Start-Up and Use To access the menu options, simultaneously press the Upper and Lower Alarm Limits buttons. Then, using the Adjust Up and Adjust Down buttons, scroll through the available menu options, as follows: #1: Trend Print. Sends the existing trend data to the NPB-290 serial port. #2: Trend Clear. Deletes the existing trend data. #3: Language Selection.
Start-Up and Use Menu Items 8 and beyond are for use only by qualified service technicians. Information about these options is available in the NPB-290 service manual. BATTERY OPERATION The NPB-290 has an internal battery that may be used to power the monitor during transport or when AC power is not available. A new, fully charged battery provides a minimum of 8 hours of monitoring time if there are no alarms or serial or analog output devices connected.
Start-Up and Use Note: As the battery is used and recharged over a period of time, the amount of time between the onset of the low battery alarm and the instrument shut-off may become shorter. It is recommended that the internal battery be replaced by qualified service personnel every 24 months. Caution: If the NPB-290 is to be stored for a period of 2 months or longer, remove the battery from the monitor prior to storage. Recharge the battery when it has not been charged for 2 or more months.
Start-Up and Use Sensors and Accessories section of this manual and in the sensor directions for use. The effects of electromagnetic interference on oximetry readings are discussed in the Troubleshooting and Maintenance section of this manual.
(Balnk Page)
TROUBLESHOOTING AND MAINTENANCE Troubleshooting EMI Interference Obtaining Technical Assistance Returning the NPB-290 Maintenance TROUBLESHOOTING WARNING: if you are uncertain about the accuracy of any measurement, check the patient’s vital signs by alternate means; then make sure the monitor is functioning correctly. WARNING: The cover should be removed only by qualified service personnel. There are no user-serviceable parts inside.
Troubleshooting and Maintenance • Do not use the NPB-290; contact qualified service personnel or your local Mallinckrodt representative. 3. The monitor does not sound a tone indicating successful completion of POST. • The monitor has failed the power-on self-test. Do not use the monitor. This tone not only indicates the successful completion of POST, but it confirms that the audible alarm is functional. Contact qualified service personnel or your local Mallinckrodt representative. 4.
Troubleshooting and Maintenance 5. The Pulse Search indicator lights after successful measurements have been made. • Check the patient. • Perfusion may be too low for the NPB-290 to track the pulse. Test the instrument on someone else. Change the sensor site. Try another type of sensor. • Excessive patient motion may be preventing the NPB-290 from tracking the pulse. Verify that the sensor is securely applied and replace it if necessary. Change the sensor site.
Troubleshooting and Maintenance EMI INTERFERENCE Caution: This device has been tested and found to comply with the limits for medical devices to the IEC 601-1-2:1993, EN60601-1-2:1994, Medical Device Directive 93/42/EEC. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.
Troubleshooting and Maintenance If assistance is required, contact Mallinckrodt’s Technical Services Department or your local Mallinckrodt representative. OBTAINING TECHNICAL ASSISTANCE For technical information and assistance, or to order parts or a service manual, contact Mallinckrodt’s Technical Services Department or your local Mallinckrodt representative. The service manual includes block diagrams and a parts list required by qualified personnel for servicing the NPB-290.
Troubleshooting and Maintenance The battery should be replaced at least every 24 months. Refer to the NPB-290 service manual for the battery changing procedure. If service is necessary, contact qualified service personnel or your local Mallinckrodt representative. Periodic Safety Checks It is recommended that the following checks be performed every 24 months: • Inspect the equipment for mechanical and functional damage. • Inspect the safety relevant labels for legibility.
SPECIFICATIONS Performance Electrical Environmental Conditions Environmental Protection Physical PERFORMANCE Measurement Range SpO2 0–100% Pulse Rate 20–250 beats per minute (bpm) Accuracy SpO2 Adults 70–100% 0–69% ±2 digits unspecified Neonates 70–100% 0–69% ±3 digits unspecified Note: Accuracies are expressed as plus or minus “X” digits (oxygen saturation percentage points) between saturations of 70–100%.
Specifications Pulse Rate 20–250 bpm Note: ±3 bpm Pulse Rate accuracy is expressed as ±3 bpm across the display range. This variation equals ± one standard deviation (1SD), which encompasses 68% of the population. Pulse Rate Display Update Frequency The Pulse Rate Display updates in less than 2.5 seconds with a 1-second change in SpO2-derived pulse rate from 30 pulses per minute (ppm) to 200 ppm.
Specifications Battery Capacity A minimum of 8 hours with a new, fully-charged battery under the following conditions: no alarms and no analog or serial output devices attached. A completely discharged battery can be fully recharged in approximately 14 hours.
Specifications ENVIRONMENTAL PROTECTION Mallinckrodt recommends that customers or technical services personnel follow local governing ordinances and recycling instructions regarding disposal or recycling of battery or other device components. PHYSICAL Weight 5.5 lbs. 2.5 kg Size 3.3 in. H x 10.4 in. W x 6 .8 in. D 8.4 cm x 26.4 cm x 17.3 cm Equipment Classification (IEC 601-1 / CSA 601.
APPENDIX A: PRINCIPLES OF OPERATION Oximetry Overview OXIMETRY OVERVIEW Pulse oximetry is based on two principles: that oxyhemoglobin and deoxyhemoglobin differ in their absorption of red and infrared light (i.e., spectrophotometry), and that the volume of arterial blood in tissue (and hence, light absorption by that blood) changes during the pulse (i.e., plethysmography).
Appendix A: Principles of Operation determine SpO2. This resistor is read when the monitor is turned on, periodically thereafter, and each time a new sensor is connected. Additionally, to compensate for differences in tissue thickness, the intensity of the sensor’s LEDs are adjusted automatically. Functional versus Fractional Saturation This monitor measures functional saturation — oxygenated hemoglobin expressed as a percentage of the hemoglobin that can transport oxygen.
Appendix A: Principles of Operation Saturation (%) 100 50 pH Temperature PCO2 2,3-DPG Fetal Hb pH Temperature PCO2 2,3-DPG Fetal Hb 0 50 PO2 (mmHg) 100 Figure 8: Oxyhemoglobin Dissociation Curve 59
(Blank Page)
APPENDIX B: Data Port Protocol Overview Trend Data Connecting to the Data Port Real-Time Display Format Patient Data and Operating Status Nurse Call OVERVIEW Patient data can be obtained through the data port on the back of the NPB-290 by connecting it to an attached PC or serial printer. When connecting the NPB-290 to a printer or PC, verify proper operation before clinical use. Both the NPB-290 and the printer or PC must be connected to a grounded AC outlet.
Appendix B: Data Port Protocol #2: Trend Clear. Clears the existing trend data, but does not erase the information. Note: To activate, press the Upper Alarm Limit button. #3: Language Selection. Provides the opportunity to change the language to one of the following: Option 0: Option 1: Option 2: Option 3: Option 4: Option 5: Option 6: English French German Italian Spanish Dutch Portuguese #4: Baud Rate. Allows the interconnection to various printers. #5: EPP Mode.
Appendix B: Data Port Protocol compatible pins (AMP part number 66570-2). The cable should be no more than 25 feet (7.6 meters) in length. The external ITE (Information Technology Equipment) device must be certified to UL-1950 or IEC-60950. The cable used must have a braided shield providing 100% coverage, such as a Belden cable (Belden part number 9609) or equivalent.
Appendix B: Data Port Protocol 9 10 11 12 13 14 15 1 2 3 4 5 6 7 8 Figure 9: Serial Port Pin Layout Pins 2, 3, and 5 provide data in RS-232 format. Pins 1, 4, 9, and 12 provide data in RS-422 format. TxD+ and TxD- are the differential transmit data pair. RxD+ and RxD- are the differential receive pair. No hardware flow control is used. However, XON/XOFF flow control is supported. REAL-TIME DISPLAY FORMAT Real-time data is continuously sent to the data port on the back of the NPB-290.
Appendix B: Data Port Protocol NPB-290 VERSION 1.0.0.
Appendix B: Data Port Protocol Software Revision Level NPB-290 TIME VERSION 1.0.0.1 %SPO2 CRC: XXXX BPM SpO2 Limit: 30-100% PA Status PR Limit: 100-180BPM The next data field tells the user the software level, (Version 1.0.0.1) and a software verification number (CRC: XXXX). Neither of these numbers should change during normal operation. The numbers may change if the monitor is serviced and receives a software upgrade. Alarm Limits NPB-290 TIME VERSION 1.0.0.
Appendix B: Data Port Protocol Patient Data NPB-290 VERSION 1.0.0.1 CRC: XXXX TIME %SPO2 BPM 12-NOV-97 14:00:05 100 190* SpO2 Limit: 30-100% PA Status 50 PR Limit: 100-180BPM Patient data is highlighted in the display above. Parameter values are displayed directly beneath the heading for each parameter. In this example the %SpO2 is 100, and the pulse rate is 190 beats per minute.
Appendix B: Data Port Protocol Table 3: Status Codes Note: Code Meaning AO Alarm Off AS Alarm Silence LB Low Battery LP Loss of Pulse MO Patient Motion PH Pulse Rate High Limit Alarm PL Pulse Rate Low Limit Alarm PS Pulse Search SH Sat High Limit Alarm SL Sat Low Limit Alarm SD Sensor Disconnect A sensor disconnect will also cause three dashes (- - - ) to be displayed in the patient data section of the printout.
Appendix B: Data Port Protocol NPB-290 VERSION 1.0.0.1 TREND TIME %SPO2 BPM 22-NOV-97 14:00:05 100 120 22-NOV-97 14:00:07 100 121 22-NOV-97 14:00:09 100 120 Output Complete SpO2 Limit: 30-100% PA 150 154 150 PR Limit: 100-180BPM Figure 11: Trend Data Printout Once a trend printout has begun, the NPB-290 must be turned off and back on again before a new trend printout can begin. A trend printout cannot be aborted without turning off the NPB-290 or the printer.
Appendix B: Data Port Protocol If battery power is disconnected or depleted, trend data and user settings will be lost. All data is stored with error detection coding. If data stored in memory is found to be corrupted, it is discarded. NURSE CALL The NPB-290 provides two different types of nurse call interfaces; an RS-232 level and relay closure. The RS-232 level nurse call function only operates when the monitor is connected to AC power. It will not operate when running on battery.
Appendix B: Data Port Protocol Table 4: Analog Pinouts Pin Parameter Parameter Range 6 %SpO2 0 - 100 13 Pulse rate 0 - 250 14 Pleth wave 0 - 255 For example, as the current value of %SpO2 varies from 0 to 100%, the voltage from pin 6 to ground (pin 10) would vary from 0 to 1 volt. A voltage of .94 volts indicates a current %SpO2 value of 94.