Instruction Manual

5. SOFTWARE OPERATION
5.1. Data logger Setup
Click on the icon on the menu bar. The Setup window will appear as
shown below; descriptions for each field in the Setup windoware listed
directly below for illustration:
The Sampling Setup field instructs the DATA LOGGER to log readings at
a specific rate. You can input specific sampling rate data at the left
Combo box and select the time unit at the right Combo box.
The LED Flash Cycle Setup field can be set 10s/20s/30s by the user
depending on the requirement. By selecting "No Light" option, there will
be no flash there by increasing the battery life.
The Alarm Setup field allows the user to set HIGH and LOW
temperature limits.
There are two start methods at the Start Method field:
1. Manual: Choose this item, the user need to click the logger button to
start data logging.
2. Automatic: Choose this item the logger will start data logging
automatic after the delay time.
User can set a specific delay time, if the delay time is O second, the
logger will start logging immediately.
Click on the SETUP button to save changes. Press the DEFAULT button
to set the Logger to factory default condition. Press the CANCEL button
to abort the setup.
Notes: All stored data will be permanently erased when Setup is
finished. To enable you to save the data before it is lost, click Cancel
and then you need to download data.
The battery might be exhausted before logger finish specified sample
points. Always ensure that the remaining power in the battery is
sufficient for completing your logging task. If in doubt, we recommend
that you always install a fresh battery before logging critical data.
7