3. LED STATUS GUIDE
REC ALM Both LED lights OFF Start logging replace battery
Logging not active and download the data
Or Low Battery
REC ALM One green flash every To start, hold the start button
10 sec.* Logging, until Green flash 4 times
no alarm condition**
Green double flash
every 10 sec.*
Delayed start
REC ALM Red double flash every Data logging, it will stop
30 sec. * -Logging, automatically. No data will be lost.
low Temperature Replace battery and download data
alarm. Red Triple flash
every 30 sec. *
-Logging, high
temperature alarm.
Red single flash every
20 sec.
-Low Battery****
REC ALM Red single flash every It won't logging until the Type-K
2 sec. -Type-K not probe connect to the logger.
connect to the logger
REC ALM Red and green single Download data
flash every 60 sec.
-Logger memory full