Operator`s manual
Measurement Screens (Parameters and Limits) (AVM-510A)
10-8
Parameter Range Absolute
Accuracy
Relative
Accuracy
Notes Min
Setting
Max
Setting
Default
Min
Default
Max
Red ampl
Nominal ± 50%
Nominal ±
±±
± 50%
± 1 IRE
±
±±
± 7 mV
± 1/2 IRE
±
±±
± 3 mV
88.2 IRE nominal
662 mV nominal
44 IRE
533 mV
148.9 IRE
985 mV
78.0 IRE
582 mV
98 IRE
936 mV
Red phase
Nominal ± 90°± 1°± 1/2° 103.4° nominal 112.9° 112.9° 98.4° 108.4°
Reference
ampl
280 to 780 mV
±
±±
± 7 mV ±
±±
± 7 mV
420 mV nominal 599 mV 599 mV 400 mV 440 mV
SC/H phase
(1)
± 90°± 5°± 5° 0° nominal -24° 24° -10° 10°
Setup
± 15 IRE ± 1 IRE ± 1/2°
7.5 IRE, NTSC
only
0.3 IRE 14.8 IRE 6 IRE 9 IRE
Sync ampl
(1)
20 to 80 IRE
150 to 600 mV
± 1 IRE
±
±±
± 7 mV
± 1/2°
±
±±
± 3 mV
40 IRE nominal
300 mV nominal
32.2 IRE
201 mV
47.7 IRE
399 mV
35 IRE
265 mV
45 IRE
335 mV
VITC Result:hh:mm:ss
White bar
ampl
50 to 138 IRE
350 to 980 mV
± 1 IRE
±
±±
± 7 mV
±
±±
± 7 mV
± 1/2°
±
±±
± 3 mV
±
±±
± 3 mV
100 IRE nominal
700 mV (color)
560 mV (freq)
-49.5%
651 mV
421 mV
49.5%
749 mV
699 mV
70 IRE
680 mV
505 mV
110 IRE
720 mV
615 mV
Yellow ampl
Nominal ± 50%
Nominal ±
±±
± 50%
± 1 IRE
±
±±
± 7 mV
± 1/2°
±
±±
± 3 mV
62.2 IRE nominal
349 mV nominal
42.5 IRE
727 mV
81.5 IRE
720 mV
52 IRE
398 mV
72 IRE
678 mV
Yellow phase
Nominal ± 90°± 1°± 1/2° 167.1° nominal 157.6° 176.6° 162.1° 172.1°
Zero
reference
pulse
50 to 138 IRE
± 1 IRE ± 1/2 IRE
On selected line
(1)
Available with Picture Guard
Range – The highest and lowest value that can be measured. Bold ranges and accuracies are for PAL automeasurement screens, when these
are different from NTSC.
Absolute Accuracy – The relationship between the value measured and the actual value.
Relative Accuracy – The relationship between the measured value in one instance and the measured value in another instance, when the actual
values are the same.
Min Setting – The lowest possible value of the minimum limit.
Max Setting – The highest possible value of the maximum limit.
Default Min/Max – Limit settings as shipped, or after a full system reset.