Operator`s manual

Measurement Screens (Parameters and Limits) (AVM-510A)
10-8
Parameter Range Absolute
Accuracy
Relative
Accuracy
Notes Min
Setting
Max
Setting
Default
Min
Default
Max
Red ampl
Nominal ± 50%
Nominal ±
±±
± 50%
± 1 IRE
±
±±
± 7 mV
± 1/2 IRE
±
±±
± 3 mV
88.2 IRE nominal
662 mV nominal
44 IRE
533 mV
148.9 IRE
985 mV
78.0 IRE
582 mV
98 IRE
936 mV
Red phase
Nominal ± 90°± 1°± 1/2° 103.4° nominal 112.9° 112.9° 98.4° 108.4°
Reference
ampl
280 to 780 mV
±
±±
± 7 mV ±
±±
± 7 mV
420 mV nominal 599 mV 599 mV 400 mV 440 mV
SC/H phase
(1)
± 90°± 5°± 5° 0° nominal -24° 24° -10° 10°
Setup
± 15 IRE ± 1 IRE ± 1/2°
7.5 IRE, NTSC
only
0.3 IRE 14.8 IRE 6 IRE 9 IRE
Sync ampl
(1)
20 to 80 IRE
150 to 600 mV
± 1 IRE
±
±±
± 7 mV
± 1/2°
±
±±
± 3 mV
40 IRE nominal
300 mV nominal
32.2 IRE
201 mV
47.7 IRE
399 mV
35 IRE
265 mV
45 IRE
335 mV
VITC Result:hh:mm:ss
White bar
ampl
50 to 138 IRE
350 to 980 mV
± 1 IRE
±
±±
± 7 mV
±
±±
± 7 mV
± 1/2°
±
±±
± 3 mV
±
±±
± 3 mV
100 IRE nominal
700 mV (color)
560 mV (freq)
-49.5%
651 mV
421 mV
49.5%
749 mV
699 mV
70 IRE
680 mV
505 mV
110 IRE
720 mV
615 mV
Yellow ampl
Nominal ± 50%
Nominal ±
±±
± 50%
± 1 IRE
±
±±
± 7 mV
± 1/2°
±
±±
± 3 mV
62.2 IRE nominal
349 mV nominal
42.5 IRE
727 mV
81.5 IRE
720 mV
52 IRE
398 mV
72 IRE
678 mV
Yellow phase
Nominal ± 90°± 1°± 1/2° 167.1° nominal 157.6° 176.6° 162.1° 172.1°
Zero
reference
pulse
50 to 138 IRE
± 1 IRE ± 1/2 IRE
On selected line
(1)
Available with Picture Guard
Range – The highest and lowest value that can be measured. Bold ranges and accuracies are for PAL automeasurement screens, when these
are different from NTSC.
Absolute AccuracyThe relationship between the value measured and the actual value.
Relative Accuracy – The relationship between the measured value in one instance and the measured value in another instance, when the actual
values are the same.
Min Setting – The lowest possible value of the minimum limit.
Max Setting – The highest possible value of the maximum limit.
Default Min/Max – Limit settings as shipped, or after a full system reset.