MM-410 Waveform/Vector Rasterizing Monitor AVM-510A Series Automated Video Monitors Operator’s Manual 946-5500-00 November 2001
MAGNI is a registered trademark of Magni Systems, Inc. Magni Systems, MagniCoder, MagniPort, PictureGuard, and Flicker Stabilizer are trademarks of Magni Systems, Inc. IBM and IBM PC AT are registered trademarks of International Business Corporation. MS-DOS and Windows are trademarks of Microsoft Corporation. All other brand or product names or systems mentioned herein are trademarks or registered trademarks of their respective manufacturers. Copyright 2001 by Magni Systems, Inc. All rights reserved.
Contents 1. About This Manual Introduction ....................................................................................................................................... 1-1 Audience............................................................................................................................................ 1-1 Organization ...................................................................................................................................... 1-1 Conventions ...........
Contents 4. Installation Introduction......................................................................................................................................... 4-1 Installing the MM-410/AVM-510A into a Rack .............................................................................. 4-1 Connecting Signal Input and Video Output Lines ............................................................................ 4-2 Connecting Auxiliaries .......................................................
Contents SPLIT SCREEN..................................................................................................................................6-4 WFM LIMIT LINE.............................................................................................................................6-4 OUT OF LIMIT LINE........................................................................................................................6-4 UPPER LIMIT LINE........................................................
Contents H BLANKING LEVEL ................................................................................................................... 6-11 COLOR SETUP ............................................................................................................................... 6-12 MEASUREMENT LINE................................................................................................................. 6-12 SIGNAL ID (NTSC) ...................................................................
Contents Using Intensity and Persistence........................................................................................................6-16 Identifying Waveforms......................................................................................................................6-17 Displaying SC/H Phase and Time Codes.........................................................................................6-18 Videotaping Waveforms...................................................................
Contents 8. Audio (AVM-510A) Introduction......................................................................................................................................... 8-1 Audio Reference Levels ..................................................................................................................... 8-2 Audio Meter ........................................................................................................................................ 8-3 Audio Limit Reporting.........
Contents 10. Measurement Screens (Parameters and Limits) (AVM-510A) Introduction........................................................................................................................................10-1 Where Parameters are Measured (Composite)................................................................................10-2 EBU/SMPTE Component Measurements (A/B Inputs).................................................................10-4 Audio............................................
Contents Audio (AVM-510A) .........................................................................................................................A-5 Interfaces (MM-410) ........................................................................................................................A-5 Interfaces (AVM-510A) ...................................................................................................................A-5 Logbook Software Specifications (AVM-510A) ................................
Contents F. Glossary ofTerms .......................................................................................................................F-1 G. Applications ................................................................................................................................. G-1 Figures 2-1. MM-410 Front View................................................................................................................. 2-3 2-2. MM-410 Rear View........................................
Contents 7-22. Two Composite Waveforms (Parade)...................................................................................7-20 7-23. Two Composite Vectors ........................................................................................................7-21 7-24. Safe Title Area.......................................................................................................................7-23 7-25. Reduced Size Waveform Display...........................................................
1 1. About This Manual Introduction This manual describes how to install, configure, and operate the MM-410 Waveform/Vector Rasterizing Monitor (MM-410) and the AVM-510A Series Automated Video Monitors (AVM-510A). The information presented will help to familiarize you with all aspects of the MM-410/AVM-510A and provide a reference tool for questions you might have in the future.
About This Manual • Chapter 11: AVM-510A Option T (For Transmission Monitoring)— Describes the functions added to the AVM-510A using the option T. • Chapter 12: AVM-510A-C (For Component Measurements)— Describes the component measurement set and the digital measurement set when an SDM-550 is interconnected. • Appendix A: Specifications— Lists the MM-410/AVM-510A electrical and mechanical specifications.
About This Manual Safety Precautions Adhere to the following safety precautions to ensure proper use of the MM-410/AVM-510A. WARNING: FOR PLUGGABLE EQUIPMENT, THE SOCKET OUTLET SHALL BE INSTALLED NEAR THE EQUIPMENT AND SHALL BE EASILY ACCESSIBLE. CAUTION: DANGER OF EXPLOSION IF BATTERY IS INCORRECTLY REPLACED. REPLACE ONLY WITH THE SAME OR EQUIVALENT TYPE RECOMMENDED BY THE MANUFACTURER. DISPOSE OF USED BATTERIES ACCORDING TO THE MANUFACTURER’S INSTRUCTIONS.
About This Manual Notes 1-4
2 2. Product Overview (MM-410) MM-410 Waveform/Vector Rasterizing Monitor The MM-410 Series monitors are dual standard (NTSC and PAL) and multi-format (Beta, MII, SVideo, EBU, GBR). Separate A and B input selections provide for easy transitions between composite and component monitoring.
Product Overview (MM-410) MM-410 Features 2-2 • Works in all standard 525 and 625 analog formats: NTSC, PAL, SMPTE, EBU, Beta, MII, S-video, GBR (RGB) • Automatically switches between 525 and 625 • Waveform and vectors can be colored separately • Displays waveform and vector together in different colors • No distortion of display due to cathode ray tube aging • No burn in from steady, bright displays • Waveform and vector images can be captured and stored • Variable persistence as well as
Product Overview (MM-410) MM-410 Front View The following LEDs, indicators, buttons, switches (knobs) can be viewed from the front of the MM-410 (see Figure 2-1). NOTE: Some buttons on the front panel have two functions when used in conjunction with the Blue (2nd) button.
Product Overview (MM-410) Power Switch Powers the MM-410 on and off. Mode Buttons The Mode buttons and the Blue button determine what the monitor displays: Waveform, Vector, Picture, or Memory Recall. Refer to Chapter 5, “Front Panel Operation,” for more information about configuring the MM-410 using the front panel buttons. Input Button This button lets you choose three signal inputs (Input A: only, Input B: only, or both Input A: and Input B).
Product Overview (MM-410) Horizontal Position Knob, LEDs, and Buttons The horizontal section of the front panel includes six buttons and one knob: Line Select (Line Sel) and Field Select (Field Sel), Sweep (Sweep) and Horizontal Magnification (H Mag), Vector Phase (Vec. Phase), and Variable Gain (Var. Gain), and a horizontal positioning knob for horizontal control options. Sweep LEDs Three Sweep LEDs (1H, 2H, and 2F) light in conjunction with the three Sweep button selections.
Product Overview (MM-410) MM-410 Rear View The following cable connectors can be viewed from the rear of the MM-410 (see Figure 2-2). Legend Connector Function 1 RS-232 Lets you connect to a PC for system updates. 2 Picture monitor outputs These outputs (S-video and composite) are available for a picture monitor display of a picture, waveforms, or signal status. 3 Composite (A) One switchable loop-through composite input.
3 3. Product Overview (AVM-510A) AVM-510A Series Automated Video Monitors The Magni AVM-510A Series Automated Video Monitors (AVM-510A) are full-featured rasterizing waveform monitors and vectorscopes with comprehensive measurement and error reporting capabilities. Magni’s patented raster display technology provides bright and sharp waveform and vector displays that resemble familiar CRT displays. All displays are routable so they can be viewed anywhere in a facility and on existing picture monitors.
Product Overview (AVM-510A) AVM-510A Features 3-2 • Works in all standard 525 and 625 analog formats: NTSC, PAL, SMPTE, EBU, Beta, MII, S-video, GBR (RGB) • Automatically switches between 525 and 625 • Waveform and vectors can be colored separately • Displays waveform and vector together in different colors • No distortion of display due to cathode ray tube aging • No burn in from steady, bright displays • Waveform and vector images can be captured and stored • Variable persistence as well
Product Overview (AVM-510A) AVM-510A Front View The following LEDs, indicators, buttons, switches (knobs) can be viewed from the front of the AVM-510A (see Figure 3-1). NOTE: Some buttons on the front panel have two functions when used in conjunction with the Blue (2nd) button.
Product Overview (AVM-510A) Power Switch Powers the AVM-510A on and off. Mode Buttons The Mode buttons and the Blue button determine what the monitor displays: Wfm, Mix,Vector, Picture, AutoMeas (Automeasure), or Memory Recall. Refer to Chapter 5, “Front Panel Operation,” for more information about configuring the AVM-510A using the front panel buttons. Input Button This button lets you choose three signal inputs (Input A: only, Input B: only, or both Input A: and Input B:).
Product Overview (AVM-510A) Position LEDs Three Position LEDs (1, 2, and 3) light in conjunction with the three Position button selections. Filter LEDs Three Filter LEDs (Luma, Chroma, and Parade) light in conjunction with the three Filter button selections. Refer to Chapter 5, “Front Panel Operation,” for more information about configuring the AVM-510A using the front panel buttons.
Product Overview (AVM-510A) AVM-510A Rear View The following cable connectors can be viewed from the rear of the AVM-510A (see Figure 3-2). Legend Connector Function 1 RS-232 computer interface A bi-directional communication interface that lets you control the AVM-510A by way of ASCII control codes. 2 Picture monitor outputs These outputs (S-video and composite) are available for a picture monitor display of a picture, waveforms, or signal status.
4 4. Installation Introduction This chapter describes how to install the MM-410 Waveform/Vector Rasterizing Monitor (MM-410) and the AVM-510A Series Automated Video Monitors (AVM-510A) and into a rack, how to connect cables to them, and how to configure them for your specific needs. Installing the MM-410/AVM-510A into a Rack The MM-410/AVM-510A mounts into any industry-standard rack. To install either one: 1. Install the rack ears onto the MM-410/AVM-510A. 2.
Installation AVM-003 Figure 4-1. MM-410/AVM-510A Rack Installation (AVM-510A Shown) Connecting Signal Input and Video Output Lines Refer to Chapter 2, “Product Overview (MM-410) and Chapter 3, “Product Overview (AVM-510A),” for detailed descriptions and the locations of the MM-410/AVM-510A rear panel signal input and video output connectors.
Installation DC Control This 9-pin connector lets you select any of the first four front panel memory locations by grounding (earthing) any one of the four pins (DC Control). An indication of the selection of one of the first three memories is given by the grounding of three other pins (refer to the following table).
Installation RS-232 Computer (AVM-510A) This is used for a direct connection to a computer. The computer can be running Logbook Software or communication software to control the AVM-510A from a remote location. RS-232 Modem (AVM-510A) This is used for a direct connection to a modem or printer. When connected to a modem, the AVM510A can be controlled over telephone lines using a computer running Logbook Software. When connected to a printer, waveforms and vectors can be printed to paper.
Installation 8. Press the Enter button to store that memory. 9. Press the Clr Mnu button to exit. 10. If you intend to view signals of several different formats, save them in other available memory locations using steps 1 through 8. NOTE: Settings for other formats or standards are also stored. For example, (1) the +V SWITCH ON/OFF setting for PAL even when memory is stored for NTSC and (2) limit settings for both standards are stored.
Installation Notes 4-6
5 5. Front Panel Operation Introduction This chapter explains how to operate the MM-410 Waveform/Vector Rasterizing Monitor (MM-410) and the AVM-510A Series Automated Video Monitors (AVM-510A) by using its front panel buttons and rotary switches (knobs). The MM-410/AVM-510A front panel is made up of four main sections (see Figures 5-1 and 5-2): • Mode • Vertical • Horizontal • Menu The remainder of this chapter describes the operational characteristics of each one.
Front Panel Operation MODE MM-410 Waveform Picture Input A Wavefront/Vector Rasterizing Monitor B 1 Vector 2 4 5 3 Display Single Alternate Memory Recall Parade 6 VERTICAL 1 Vertical Position Position V Mag Filter Split Screen 2 3 Luma Chroma Parade MM-410 MODE Waveform Picture VERTICAL 1 Position Input A Wavefront/Vector Rasterizing Monitor Vertical Position B 1 Vector 2 3 Display Single Alternate Memory Recall 4 5 6 Parade Horizontal Position 3 Luma 1H V Mag 2 Fi
Front Panel Operation MODE AVM-510A Mix Automated Video Monitor Wfm 1 A Picture Input 2 B 3 Single Vector Memory Recall 4 Auto Meas 5 Display Alternate 6 Parade VERTICAL 1 Vertical Position 2 3 Wfm Pos Mag Vector Mag Luma Chroma Split Screen Filter Parade MODE AVM-510A VERTICAL Wfm 1 A Picture 2 Input B Single Vector Memory Recall 4 Auto Meas 5 Vertical Position 2 3 3 Display Alternate 6 Parade MENU HORIZONTAL 1 Mix Automated Video Monitor 1H Wfm Pos Mag F
Front Panel Operation Monitor Display Explanation Figure 5-3 shows what you can see on a monitor display as you operate the MM-410/AVM-510A using the front panel buttons and knobs. The table that follows explains each of the display areas. Figure 5-3.
Front Panel Operation Figure Legend Description 1 Indicates the selected and displayed input signal A or B, the displayed signal system 525 (NTSC) or 625 (PAL), and the displayed signal format digital for the A or B input. 2 Field and Line – Field and line number display. 3 Waveform – Waveform display. 4 Audio Meters – Analog audio metering. 5 Picture – Picture display. 6 Vector – Vector display. 7 Gain – Variable gain display (waveform/vector).
Front Panel Operation Vertical These buttons let you manipulate and vertically position waveforms on a display monitor. The following table lists the functions of the Vertical buttons and knob. Button/Knob Function Vertical Position Positions the waveform on a display monitor. Use this knob to move the waveform up or down. Also moves the three parts of a component waveform separately, depending on the state of the Position button and its associated LEDs (Wfm Pos for the AVM-510A).
Front Panel Operation Horizontal These buttons let you manipulate and horizontally position waveforms on a display monitor. The following table lists the functions of the Horizontal buttons and knob. Button/Knob Function Horizontal Position Positions the waveform on a display monitor. Use this knob to move the waveform left or right. The Horizontal position knob also has a scrolling function when used in magnified sweep or vector phase mode.
Front Panel Operation Button/Knob Function Var. Gain Reduces a signal input (waveform and vector) by a size factor of 0.7 or increases a signal input (waveform and vector) by a size factor of approximately 1:2. When selected, the corresponding LED lights. You can also use Var (Variable) Gain with V Mag if you need to see a more detailed waveform or vector. When selected, the word GAIN appears at the top-right of the display. Note: If you have both Line Sel and Vec.
Front Panel Operation Measurement menu options include: • SELECT SET • RESPONSE TIME • NOISE LINE • TIME READOUT • REPORT • WAVEFORM READOUTS • AUDIO OPTIONS • MULTIBURST ID Refer to Chapter 6, “Operating the MM-410/AVM-510A,” for more information about the Config and Mesmnt menu options. How the Menu Works Following are some basic rules about how the menu works. • You do not have to exit one menu to access another.
Front Panel Operation • Any menu option selected takes effect immediately. You must de-select a menu option to erase it. • A blue, highlighted cursor indicates the place in the menu structure. You can move the cursor using the Up and Down arrow buttons. • Menu entries in all capital letters indicate that there are sub-menus below that entry. • The Blue button is used for quick access to the functions available for the MM-410/AVM-510A.
Front Panel Operation You can exit a menu using several methods: • Pressing the Prev Mnu button closes the current menu and moves you back to the previous menu. If you press it enough times, you will exit all menus. • Pressing the Clr Mnu button closes all menus immediately. • If you are down several layers in the menus, pressing the Config button once brings you to the top Configure menu. Pressing it again closes the Configure menu. Most menu options that are selected take effect immediately.
Front Panel Operation Quick Access to Commonly Used Functions The Blue button provides quick access to certain functions, such as status recall. Note that any Blue button function is also available through the main menus. The options that the Blue button can control are the ones in blue boxes on the front panel. To execute any option labeled in a blue box, press the Blue button and then the button you want. The following table lists the available options. Button Function Waveform Recall memory 1.
Front Panel Operation Factory Reset To reset the MM-410/AVM-510A back to its factory default settings: 1. Quickly press the Blue button twice. Afterwards the following message appears on the monitor screen: Press Enter for FACTORY RESET 2. Press the Enter button to reset or any other key to cancel the reset. If you do not press the Enter button after a period of time, the reset is automatically canceled.
Front Panel Operation Notes 5-14
6 6. Operating the MM-410/AVM-510A Introduction This chapter describes how to operate the MM-410 Waveform/Vector Rasterizing Monitor (MM-410) and the AVM-510A Series Automated Video Monitors (AVM-510A). By now you should have the MM-410/AVM-510A installed and configured for your specific needs. Using the information provided in this chapter, you can now select, monitor, automeasure, and report the appropriate display options available in the Config (Configure) and Mesmnt (Measurement) menus.
Operating the MM-410/AVM-510A Config Menu The Config menu contains the display portion of the MM-410/AVM-510A. It lets you view the input, display, vector, memory, time, alarm settings, and serial port options shown in Figure 6-1. The following sections describe each of the Config menu options.
Operating the MM410/AVM-510A INPUT Sub-Menu This section describes each of the INPUT sub-menu options. Input A: and Input B: lets you select a displayed input and the format of the selected input. INPUT A: Lets you select either a CPST-1 (Composite) or S-video signal input for display when Input A: is selected. INPUT B: Lets you select the CPST-2 (Composite), SMPTE/EBU, BETA, MII/S, S-video, or GBR signal input for display when Input B: is selected.
Operating the MM-410/AVM-510A LINE SELECT MODE Lets you select either a 2 line or 15 line display. SAFE AREA Lets you turn on/off the safe title and safe action areas of a display. Safe title is limited to 90% of the display (no part of the NTSC/PAL display can be closer than 10% to the edge). Safe action is limited to 95% of the screen display (no part of the NTSC display can be closer than 5% to the edge).
Operating the MM410/AVM-510A VECTOR Sub-Menu This section describes each of the VECTOR sub-menu options. +V SWITCH Lets you turn on/off the +V switch for a vector display. When off, a PAL type vector is displayed with 12 boxes. When On, an NTSC type vector is displayed with 6 boxes.
Operating the MM-410/AVM-510A PRINTER SETUP Lets you select and display the following options: PRINTER TYPE − Display the type of printer connected to a computer (HP DeskJet 500C, HP Laser Jet, HP DeskJet, or Epson FX-85. PRINT DENSITY − Lets you see the print density of a display in the form of a horizontal bar graph. BAUD RATE − Lets you select a baud rate for data transfers between the AVM-510A and a computer. Available baud rates: 9600, 19,200, 28,880, 38,400, 57,600, and 115,200.
Operating the MM410/AVM-510A SET INTERNAL CLOCK Lets you set the internal clock of the MM-410/AVM-510A. The selections are: YEAR, MONTH, DAY, HOURS, and MINUTES, and SECONDS. Select Press To Set and then the Enter button to save the internal clock settings. SYSTEM IDS Lets you display the MM-410/AVM-510A system software and hardware identification information for: Loader, Software, Fnt Pnl, Hardware, and Board ID. MEMORIES Sub-Menu This section describes each of the MEMORIES sub-menu options.
Operating the MM-410/AVM-510A Mesmnt Menu The Mesmnt menu contains the measurement portion of the MM-410/AVM-510A (see Figure 6-2 and Figure 6-3). The Mesmnt menu lets you view the A and B input (digital) measurement options. The following sections describe each of the Mesmnt menu options.
Operating the MM410/AVM-510A Measurement (Composite PAL, one each for Input A and Input B) AUDIO OPTIONS SELECT SET NOISE LINE, 10 to 35 SYSTEM VIDEO RESPONSE TIME MEASURE LINE, 1/3:10 to 2/4:262 (bar graph) SIGNAL ID, AUTO SIG ID/LIVE VIDEO/75% COLOUR BARS/ AUDIO METER, ON/OFF 100% COLOUR BARS/MULTIBURST/CCIR 17/330/ REF LEVEL dB, -20 to 10 NATIONAL ITS/CCIR18 AUDIO REPORT SYS VIDEO LIMITS REPORTING, ON/OFF SYNC AMPL HIGH AUDIO LIMIT BURST AMPL HIGH REPORT LIMIT, -20 to 10 dB PEAK VIDEO HIGH ALARM DELAY
Operating the MM-410/AVM-510A TIME READOUT Sub-Menu The TIME READOUT sub-menu options let you select the TIME, TIME SOURCE, or VITC LINE options. TIME Lets you turn ON/OFF the internal clock. TIME SOURCE Lets you display one of the following time sources: VITC TIMECODE or INTERNAL CLOCK. The TIME SOURCE option also lets you automatically switch between VITC TIMECODE or INTERNAL CLOCK by selecting AUTO SWITCH. VITC LINE Lets you select a VITC line between 9 and 25.
Operating the MM410/AVM-510A SELECT SET Sub-Menu (AVM-510A, Composite NTSC) The SELECT SET sub-menu options let you select SYSTEM VIDEO, COLOR SETUP, FREQUENCY RESP, NON LINEAR, or AUTOMEASURE (see Figure 6-2). When AUTOMEASURE is selected the AVM-510A automatically enables the proper measurement set for the line selected in the signal under test. The selected set appears in the upper-left corner of the AUTOMEASURE display screen.
Operating the MM-410/AVM-510A COLOR SETUP Lets you select the MEASURE LINE, SIGNAL ID, or COLOR LIMITS options (COLOUR LIMITS for PAL). MEASUREMENT LINE Lets you select a line (1/3:10 to 2/4:262) that measurements can be taken from (1/3:7 to 2/4:612 for PAL). SIGNAL ID (NTSC) Lets you select LIVE VIDEO, 75% COLOR BARS, 100% COLOR BARS, 75% SMPTE BARS, FCC MULTIBURST, FCC COMPOSITE, NTC7 COMPOSITE, MULTIBURST or AUTO SIG ID.
Operating the MM410/AVM-510A SIGNAL ID (PAL) Lets you select AUTO SIG ID, LIVE VIDEO, 75% COLOUR BARS, 100% COLOUR BARS, MULTIBURST, CCIR 17/330, NATIONAL ITS, or CCIR18. When AUTO SIG ID is selected, the AVM-510A automatically detects from the selected line what type of test signal is present. The line number and type of test signal present appears in the upper-right corner of the AUTO MEAS SET display screen. FREQ RESP LIMITS Lets you select and set frequency response limits.
Operating the MM-410/AVM-510A AUTOMEASURE (NTSC) Lets you select a line (1/3:10 to 2/4:262) that measurements can be taken from. Refer to Appendix F, “Options (AVM-510A),” for more information about optional measurement sets. AUTO MEAS SET (PAL) Lets you select a line (1/3:7 to 2/4:612) that measurements can be taken from. RESPONSE TIME (AVM-510A) Lets you select the response time via a bar graph. NOISE LINE (AVM-510A) Lets you select a noise line between 10 and 35.
Operating the MM410/AVM-510A REPORT SUSTAINED ERRORS Lets you turn OFF or set the time (in minutes) of REPORT SUSTAINED ERRORS. REPORT ALL ERRORS CLEAR Lets you turn ON/OFF the REPORT ALL ERRORS CLEAR option. REPORTING DELAY Lets you turn ON/OFF the REPORTING DELAY option. AUDIO OPTIONS Sub-Menu (AVM-510A) The AUDIO OPTIONS sub-menu options let you select the AUDIO METER, REF LEVEL dB, or AUDIO REPORT options. AUDIO METER Lets you turn ON/OFF audio meter displays for the analog audio inputs.
Operating the MM-410/AVM-510A Using Intensity and Persistence Intensity and persistence are set with two controls in the DISPLAY menu. To change either one: 1. Press the Config button. 2. Select DISPLAY and then press the Enter button. 3. Select INTENSITY and then turn the knob to change it. The bar changes in length and you will see an immediate change in the waveform or vector display. 4. Select PERSISTENCE and then turn the knob to change it.
Operating the MM410/AVM-510A Identifying Waveforms If you need to measure signals in different standards (composite, SMPTE, GBR, and Beta), you can select a different waveform and vector color for each standard, and then store them in separate memories. Afterwards, when you recall a memory, the MM-410/AVM-510A switches to the proper video standard and displays the waveform in a distinctive color. To establish colors associated with standards: 1. Press the Config button. 2.
Operating the MM-410/AVM-510A To create labels, decide what you want the label to say. You get 12 characters for each label. 1. Press the Config button. 2. Select INPUT and then press the Enter button. 3. Select LABEL INPUTS and then press the Enter button. 4. Use the Arrow buttons to select A: and then press the Enter button. 5. Turn the knob to select the first character of your label. After it appears in the cursor, press the Enter button. 6.
Operating the MM410/AVM-510A To view the SC/H Phase and Color Frame: 1. Press the Mesmnt button. 2. Select WAVEFORM READOUTS and then press the Enter button. 3. Select SC/H PHASE and then press the Enter button to turn it on or off. 4. Select COLOR FRAME and then press the Enter button to turn it on or off. 5. Press the Clr Mnu button to exit. The time display in the lower-right corner shows either VITC time code, taken straight from the signal, or the internal clock.
Operating the MM-410/AVM-510A Videotaping Waveforms Because the output of the MM-410/AVM-510A is a video signal, you can handle it as you would handle any other video signal. There are two video outputs, both available at the same time: one VHS and one S-video. Note that the output is always NTSC when the input is a 525-line system, and it is always PAL when the input is 625. It is not possible to look at a problem in 625 EBU and record the output on an NTSC VCR.
Operating the MM410/AVM-510A 20 Second Delay When any control setting on the front panel is changed, the MASTER ALARM is disabled for 20 seconds. This prevents the AVM-510A from sending an alarm while it changes from one mode to another. After the 20 second delay, any change in the signal that causes an out of tolerance condition triggers the MASTER ALARM immediately.
Operating the MM-410/AVM-510A To connect at short distances, you need: • A computer running Windows 3.1, 95, 98, or NT with an available serial port • A 9-pin serial cable • Logbook Software To configure the AVM-510A for a modem connection: 1. Connect the AVM-510A using the RS-232/Modem port at the rear panel to an available modem connection (see Figure 3-2). 2. Press the Config button. 3. Select OTHER and then press the Enter button. 4. Select SERIAL PORTS and then press the Enter button. 5.
Operating the MM410/AVM-510A Printing Waveforms and Vectors (AVM-510A) You can print anything you see in an AVM-510A display except the picture. The picture display is not processed in the AVM-510A, so it is not available in digital form. Before you print, you must connect a printer and then configure the AVM-510A for the printer you want to use. Note that the output to the printer is black and white, even if the waveform is in color.
Operating the MM-410/AVM-510A Print the vector using several settings of Print Density. At low settings, you will get nothing but the vector graticle. At higher settings the dots appear, and at very high settings the transitions are visible in the print. Print a waveform of live video. With a low setting for Print Density, none of the chroma or noise is printed. Higher settings print all the noise and chorma.
7 7. Viewing Waveforms, Vectors, and Pictures Introduction This chapter shows you how to view waveforms, vectors, and pictures using the MM-410 Waveform/Vector Rasterizing Monitor (MM-410) and the AVM-510A Series Automated Video Monitors (AVM-510A).
Viewing Waveforms, Vectors, and Pictures Displaying a Waveform Pressing the Waveform button lets you see a waveform for the selected input signal. Once displayed, you can move and expand it vertically and horizontally using the Vertical/Horizontal (knobs) and V Mag/H Mag buttons. Figure 7-1 shows a basic waveform display. Figure 7-1. Basic Waveform Display NOTE: Pressing the Waveform button twice always results in a two-line display with no line select and no magnification.
Viewing Waveforms, Vectors, and Pictures Under certain situations a 2 Field sweep rate is not available. For example, if you are unable to view filters in Parade at the 2 Field sweep rate but you can see either the Luma or Chroma filter as two fields. Also, you cannot display a vector with a waveform at the 2 Field sweep rate. Figure 7-2. 1H Sweep Rate (A: 525 CPST) The following table lists the sweep selections and sweep rates for the MM-410/AVM-510A.
Viewing Waveforms, Vectors, and Pictures Waveform Limit Lines Waveform limit lines cause the waveform to turn red each time a portion of it goes outside a predetermined set of limits. Two specific limits can be set: an upper limit for peak video and a lower limit for sync tip. When an upper limit is set, the waveform turns red every place the signal exceeds the upper limit, including the vertical limit.
Viewing Waveforms, Vectors, and Pictures Setting Waveform Limit Lines Waveform limit lines show any part of the input signal that goes above or below a pre-determined set of limits. Note that these limits are shown relative to the signal blanking level, not the graticule, so moving the waveform up or down does not move the beginning of the red area (see Figure 7-4). To set waveform limit line values or to toggle this feature on and off: 1. Press the Config button to access the Configure menu. 2.
Viewing Waveforms, Vectors, and Pictures Switching PAL Graticule to Percent The PAL waveform graticule is scaled, in volts, by default (Figure 7-5). For some measurements you might want to change this option to percent (Figure 7-6). To switch the PAL graticule from volts to percent: 1. Press the Config button. 2. Select DISPLAY and then press the Enter button. 3. Select GRATICULE OPTIONS and then press the Enter button. 4. Select PAL GRATICULE. 5.
Viewing Waveforms, Vectors, and Pictures How to View a Selected Line (Waveform, Vector and CPNT Bars) You use the front panel to view a selected line from an odd field or an even field. The MM-410/AVM510A is normally set up to view all lines of the video signal, however, you can change this setting to view any two or 15 lines by pressing the Line Sel button and accessing the Configure menu. Whether you view one or two lines depends on the sweep rate.
Viewing Waveforms, Vectors, and Pictures The most efficient way to use line select is with the two-field display. When you use 2 Field sweep and line select together, the selected lines are brightened in the two-field display. Figure 7-7. Line Select Indicator Viewing a Vertical Interval You can view a vertical interval using two methods: 2F sweep (expanded horizontally) and line select (see Figures 7-8 and 7-9). In 2F sweep mode, you can view both fields at once.
Viewing Waveforms, Vectors, and Pictures 7. Press the H Mag button to reduce the two-field magnification. 8. Press the Sweep button once or twice to view a one-line or two-line waveform display. 9. To go back to viewing the entire vertical interval, press the Blue and Sweep buttons and then the H Mag button. Figure 7-8. 2F Sweep, Highlighting Lines 17 and 18 Figure 7-9.
Viewing Waveforms, Vectors, and Pictures Using Filters Filters let you view only a luminance display or a chrominance display. You can parade these waveform displays with the normal or flat waveform to view a signal in three ways at the same time. Paraded filters can also be separated vertically. NOTE: The luminance display is also available for component signals. To view the effects of a filter on a composite waveform: 1. Press the Filter button.
Viewing Waveforms, Vectors, and Pictures Figure 7-11. Chroma Filter Figure 7-12. Parade of Filters Displaying a Component Signal You can display component waveforms in a variety of ways: Y (or G) channel only, all three channels superimposed on each other, or three channels paraded (see Figures 7-13 to 7-15). The component waveform mode will be displayed in the following formats: • SMPTE/525 or EBU/625 • Beta • MII • GBR (or RGB) NOTE: GBR and RGB are the same.
Viewing Waveforms, Vectors, and Pictures All component signals must use Input B:. To configure the B input: 1. Press the Config button. 2. Select INPUT and then press the Enter button. 3. Select INPUT B:. 4. Turn the knob or press the Enter button until the desired component signal is listed. 5. Press the Clr Mnu button to exit. NOTE: In this manual S-video signals are not considered component signals. S-video signals are treated like composite signals with regard to menu selections.
Viewing Waveforms, Vectors, and Pictures Figure 7-14. Beta Signal in Alternate with B-Y and R-Y Raised Figure 7-15.
Viewing Waveforms, Vectors, and Pictures Displaying a Bowtie Signal Bowtie is a specific component video signal used to accurately measure the time delay between the components. Bowtie signals consist of continous sine waves in the three components. In the main channel (Y or G), the sine wave is one frequency. In the other two channels, a slightly higher frequency is used. The timing of the sine waves are exactly in phase at the center point.
Viewing Waveforms, Vectors, and Pictures Viewing a Vector Vectors can be viewed at any time by pressing the Vector button (see Figures 7-16 through 7-18). Pressing the Vector button twice shows a vector and a waveform together. Unless you changed it, the vector color is always yellow so it can be seen easily with a waveform. Note that the vector rose (the circular graticule for vectors) is simplified when it is displayed with a waveform. Component signals can also be displayed as vectors.
Viewing Waveforms, Vectors, and Pictures This process saves the color you just selected as well as these settings: display type, input standard, position, intensity, and whether or not safe area is on or off. Figure 7-16. Composite Vector Figure 7-17. Component Waveform and Vector Figure 7-18.
Viewing Waveforms, Vectors, and Pictures Controlling the Vector Display You use the right knob to control the vector display. This switch acts as the variable gain control for vectors as well as waveforms. The switch also has the same scrolling action as the H Pos knob for expanding waveforms. To rotate the vector display, press the Vec. Phase button and then turn the knob. The Vec. Phase LED lights indicating that the switch is now controlling the vector phase.
Viewing Waveforms, Vectors, and Pictures Figure 7-19. PAL Vector Display Figure 7-20. PAL Vector with +V Switch Locking the Vector to External Sync Vector displays can be locked to an internal or external reference. To lock a vector display to the external sync source, press the Blue button and then press the Field Sel (Ext Ref) button. Repeat this sequence to go back to the internal reference. When the MM-410/AVM-510A is externally referenced, the word EXT appears at the top center of the display.
Viewing Waveforms, Vectors, and Pictures When the vector display is externally referenced, note that the burst vector does not lie along the 180º line. You can rotate this display using the right knob or you can preset the rest position of the external sync as follows: 1. Press the Config button. 2. If REFERENCE is not EXT, press the Enter button to select it. 3. Select VECTOR and then press the Enter button. 4.
Viewing Waveforms, Vectors, and Pictures The way to compare two component waveforms or vectors is to view one of them and then save it as an image, view the other waveform or vector, and then recall the image of the first. Refer to the next section “Saving Waveform and Vector Images” for more information about saving waveform and vector images. Figure 7-21. Two Composite Waveforms (Alternate) Figure 7-22.
Viewing Waveforms, Vectors, and Pictures Figure 7-23. Two Composite Vectors Saving Waveform and Vector Images The MM-410/AVM-510A has space to save up to 6 waveform or vector images. Before saving an image, keep the following in mind: • You can save the image of a waveform or a vector or a waveform and vector together. • The graticule is not saved with the image. • You cannot move the position of the recalled image. • Pictures cannot be saved.
Viewing Waveforms, Vectors, and Pictures When you recall an image, you get what was last stored at that location or nothing (if no settings were ever stored). Use the IMAGE DENSITY control in the MEMORIES menu to change the brightness of the image. You must do this before storing it. After the image is stored and recalled, the brightness cannot be changed. Locking and Clamping the Signal The following sections describe how to lock and clamp a signal.
Viewing Waveforms, Vectors, and Pictures To change clamp speed: 1. Press the Config button. 2. Select INPUT and then press the Enter button. 3. Select CLAMP and then press the Enter button or turn the knob to change between SLOW and FAST. 4. Press Clr Mnu to exit. Viewing a Picture with Waveforms Most waveform montors do not let you view a picture.
Viewing Waveforms, Vectors, and Pictures Figure 7-25. Reduced Size Waveform Display This display offers several choices. For example, the display can show the waveform alone, the vector alone, or both. The background of the waveform/vector can be opaque, so you cannot see the picture through it, or it can be mixed with the picture (see Figure 7-26). To make these choices: 1. Press the Config button. 2. Select DISPLAY and then press the Enter button. 3. Select SPLIT SCREEN and then press the Enter button.
Viewing Waveforms, Vectors, and Pictures Figure 7-26. Picture Display with Waveform/Vector Mixed Monitoring Signal Quality with Picture Guard (AVM-510A) Picture Guard lets you watch a picture while monitoring certain signal elements or parameters. With a System Video measurement screen selected, you can separately enable and disable up to 12 measured items of video picture signal integrity.
Viewing Waveforms, Vectors, and Pictures All the parameters that are monitored by Picture Guard can be found in the System Video automeasurement screen. To view these lists (see Figure 9-1): 1. With an INPUT A: or INPUT B: selected, press the Mesmnt button. 2. Select SELECT SET and then press the Entr button. 3. Select SYSTEM VIDEO and then press the Entr button. 4. Press Clr Mnu to exit. Press the Auto Meas button to view a list of parameters that are monitored in Picture Guard.
Viewing Waveforms, Vectors, and Pictures Figure 7-27. Picture Guard Locations Picture Reference While External Reference is On The picture is normally referenced to itself when displayed, even if the MM-410/AVM-510A is set to REFERENCE (EXT). This prevents the picture from shifting up and down, left and right, if the timing of the external sync is different from the displayed signal.
Viewing Waveforms, Vectors, and Pictures Comparing Two Signals with a Picture A special picture display (A/Ref Split) combines the external reference signal with Input A in one picture, alternating the two in horizontal bands (see Figure 7-28). You can use this display to match the two images for horizontal timing, chroma gain and chroma phase. In this display the signal from the A Input and the signal on the External Reference connector are compared using a picture-like display.
Viewing Waveforms, Vectors, and Pictures Notice that the second two items appear only when they are marked ACTIVE. If they are DISABLED, they will not appear when the Split Screen button is pressed. To use this display, the video signal must be the same for both inputs. If both sources have color bars, you can do a reasonable job of adjusting by matching the signals with the same live video picture. Adjust the horizontal timing of one of the two signals to align the bands.
Viewing Waveforms, Vectors, and Pictures Notes 7-30
8 8.
Audio (AVM-510A) Audio Reference Levels The AVM-510A lets you set an audio reference level (see Figure 8-1). The audio meter indicator turns yellow above the 0 reference mark and red when the level exceeds 8 dBu above this set reference mark. The horizontal marks in the audio meter represent 10 dBu increments. Note that the dBu unit is the same as dBm. To change the audio meter reference level: 1. Press the Mesmnt button to access the Measurement menu. 2.
Audio (AVM-510A) Audio Meter Audio meters can be turned on or off using the AUDIO METER menu option. Once selected, the audio meter appears at the left and right of all monitor displays. To make changes to any of the audio meter options, follow these steps: 1. Press the Mesmnt button to access the Measurement menu. 2. Select AUDIO OPTIONS and then press the Enter button. 3. Select AUDIO METER and then press the Enter button to switch between ON or OFF. Figure 8-2 shows the AUDIO METER menu options. 4.
Audio (AVM-510A) Audio Limit Reporting You can set audio limits to generate a report, activate an alarm, or activate a Picture Guard indicator if audio levels are above or below the set limits for longer than a preset time. The report can include both audio and video and can be initiated by an audio out of limits indication. Audio is included with video reporting if enabled in the audio limit settings. To enable audio out of limits to initiate a report: 1. Turn on the audio meter.
9 9. Automatic Measurement (AVM-510A) Introduction Automeasurement uses the full power of the AVM-510A Series Automated Video Monitors (AVM510A) signal processing capabilities. Automeasurement is a flexible system that can take specific kinds of measurements on the most common types of test signals and present these measurements as graphs, as numerical values, and as a red (out of limits) or green (within limits) indications.
Automatic Measurement (AVM-510A) On the extreme left side is a series of boxes, some with a single dot, some with a “P,” and some empty. They indicate the following: • Single Dot current value being measured. • P current value being measured and monitored by Picture Guard. • Empty current value is not being measured (no graph or numerical value displayed). To change the state of a parameter line, press one of the Arrow buttons. Afterwards, one of the squares turns green.
Automatic Measurement (AVM-510A) Test Signal Measurement Sets Three basic composite test signal measurement sets are available with the AVM-510A. Two of the measurement sets are associated with common test signals, such as color bars and multiburst signals. The third measurement set is for live video. The test signal measurement sets can be applied to several test signals, for example, 75% or 100% color bars.
Automatic Measurement (AVM-510A) • Forcing Live Video The choice usually used with System Video allows all System Video parameters to be measured on a live video signal. In the two other measurement sets, forcing Live Video causes the AVM-510A to measure only those parameters that can be measured on a live video signal, such as Burst Ampli and Peak Video. The other parameters are grayed out or turned off. The measurement set can be applied to a line in the vertical interval or to an active video line.
Automatic Measurement (AVM-510A) When using AUTOMEASURE/AUTO MEAS SET mode, the AVM-510A recognizes these signals: Test Signals 525 Systems 625 Systems 75% color bars, with setup 100% color bars, with setup SMPTE color bars* 75% colour bars 100% colour bars FCC multiburst CCIR 18 FCC composite NTC7 composite CCIR 17* & CCIR 330* National ITS Note that when selecting a line number for SMPTE Color Bars, CCIR 17, and 330: • 75% SMPTE Bars is never a vertical interval signal.
Automatic Measurement (AVM-510A) To force the AVM-510A to use a specific measurement set: 1. Press the Mesmnt button. 2. Select SELECT SET and then press the Entr button. 3. Select the appropriate measurement set and then press the Entr button. This forces the AVM510A to use the selected measurement set. 4. Select MEASUREMENT LINE and then turn the knob to specify the line that you want measured.
Automatic Measurement (AVM-510A) Setting Automeasurement Limits The AUTOMEASURE/AUTO MEAS SET menu lets you set measurement limits over very wide ranges. You can set different limits for Input A: and Input B:, for 525 and 625 systems, and for different component systems. After setting the limits for any measurement set, the green (safe) area changes in the bar graph.
Automatic Measurement (AVM-510A) 9. Press the Entr button to move the cursor to the MIN column and then turn the knob to set the minimum acceptable value. 10. Press the Entr button to move the cursor to the MAX column and then turn the knob to set the maximum acceptable value. While in the MIN and MAX columns, the Arrow buttons switch the cursor back and forth between them. Pressing the Entr button moves the cursor from MIN to MAX, or from MAX out and down to the next parameter name (step 8).
Automatic Measurement (AVM-510A) Under the main MEASUREMENT menu is a sub-menu called MULTIBURST ID. This is the key to multiburst signal detection when using AUTOMEASURE/AUTO MEAS SET. Three selections under MULTIBURST ID are available: • STANDARD (FCC multiburst or CCIR 18) If the AVM-510A detects a multiburst signal of any kind, it assumes it is one of these two standards, depending on the line rate of the signal.
Automatic Measurement (AVM-510A) The first step is to find out the timing of each signal part: 1. Look at the multiburst signal as a waveform. Use the 1H sweep rate. Note that the major graticle divisions represent 5 µs increments (see Figure 9-2). Figure 9-2. 1H Sweep, Left Line 2. Use the H POS knob to move this wide, two-line display so the center sync pulse is near the left side of the screen. You should now see one line all the way across the screen with a sync pulse at the left. 3.
Automatic Measurement (AVM-510A) 4. Using only the V POS knob (do not touch the H POS knob), move the waveform up and down and then write down the timing measurements for each of the following (see Figure 9-4): – Beginning of the white bar – Beginning of the pedestal after the white bar – Beginning of each of the frequency packets Figure 9-4. Measuring the Times Once you have all the figures, continue with the second step of the proccess: 1.
Automatic Measurement (AVM-510A) 10. Select each of the packets and then turn the knob to set the frequency, in 0.1 MHz increments, to the frequency of that packet in the multiburst. Note that the frequencies shown in the automeasurement display does not change to reflect your entry here. However, the measurement system must have the correct frequency to operate properly. 11. Press the Clr Mnu button to exit.
Automatic Measurement (AVM-510A) Measurement Parameters Used with Live Video Parameters System Video Color Setup Frequency Response Non-Linear Transmitter SYNC AMPL X X X X X BURST AMPL X X X X X PEAK VIDEO X PEAK LUMA X APL X NOISE X H BLANKING X H-REF TIMING X BURST REF PHA X SC/H PHASE X COLOR FRAME X BAR AMPL X X X X X X X X WHITE BAR AMPL X BLANKING LEVEL X The other measurements (not in italic) are made on the line shown at the top of the measurement scre
Automatic Measurement (AVM-510A) To select the noise measurement line: 1. Press the Mesmnt button. 2. Select NOISE LINE and then turn the knob to select a line between 10 and 35. 3. Press the Clr Mnu button. The other measurement sets (Color Setup, Frequency Response, and Non Linear) can also sample live video when forced to do so. Depending on which measurement set has been selected, only certain ones of the parameters SYNC AMPL, BURST AMPL, PEAK LUMA, NOISE, and BAR or WHITE BAR AMPL remain active.
Automatic Measurement (AVM-510A) Reports can be sent to a PC to be displayed on its monitor screen and saved as an ASCII text file. In this configuration, the AVM-510A RS-232 computer interface is connected to one of the computer’s serial ports using a 9-pin cable. Reports can also go directly to a serial printer, and a page of the measurement is printed each time there is a full page of data. This configuration uses the AVM-510A RS-232 modem interface and a 9-pin serial cable.
Automatic Measurement (AVM-510A) Sending Reports to a PC To send measurement reports from the AVM-510A to a PC: 1. Follow the procedures previously described in “Automatically Reporting Measurements.” 2. Press the Mesmnt button. 3. Select REPORT and then press the Entr button. 4. Select ERROR REPORTING and then press the Entr button to select ON. 5. Select REPORT SUSTAINED ERRORS and then turn the knob until it shows 1 Minute. 6. Press the Prev Mnu button. 7.
Automatic Measurement (AVM-510A) In approximately 20 seconds you should see a light flashing on your printer signaling you that data is being sent from the AVM-510A. You should start to see printed a series of measurements with the title “New Error.” Every minute after this you should see printed a new list of measurements with the title “Sustained Error.” If, after 5 minutes, you do not see anything from the printer, something is set up wrong between the AVM-510A and the printer.
Automatic Measurement (AVM-510A) SCHEDULED REPORT INTERVAL Select this item and then turn the knob to set this selection for any time between 5 minutes and 480 minutes (8 hours). Reporting occurs at these intervals, starting immediately. The report ocurs whether there is an error or not. If you set this item to 0 (OFF), just below the 5 minute point, periodic reporting does not occur at all, and you will only see error reporting (if error reporting is enabled).
Automatic Measurement (AVM-510A) Special Considerations for Alarm Outputs An alarm output can be set to activate with a preset delay when there is an error in a measurement set. Both closed captioning and audio missing can also be set to activate the alarm output. Alarm Output Pin one on the DC CONTROL/ALARM connector is dedicated to an alarm.
Automatic Measurement (AVM-510A) Notes 9-20
10 10. Measurement Screens (Parameters and Limits) (AVM-510A) Introduction This chapter lists details about the AVM-510A Series Automated Video Monitors (AVM-510A) signal processing capabilities automeasurement screens. The following tables are included: • Where Parameters are Measured lists each parameter, shows which measurement set(s) a parameter is found in, and whether a parameter is measured in NTSC, PAL, both NTSC and PAL, EBU, and SMPTE.
Measurement Screens (Parameters and Limits) (AVM-510A) Where Parameters are Measured (Composite) Measured Parameter System Video Color Setup NonLinear Frequency Response 0.5 MHz reference NTSC 1.0 MHz PAL 1.25 MHz NTSC 2.0 MHz Both 3.0 MHz NTSC 3.58 MHz NTSC 4.0 MHz PAL 4.1 MHz NTSC 4.8 MHz PAL 5.
Measurement Screens (Parameters and Limits) (AVM-510A) Measured Parameter System Video Color Setup Green ampl Both Green phase Both H blanking(1) Both H reference timing(1) Both Non-Linear K-2T factor Both Lum linearity PAL Noise(1) Both P/B ratio Both Peak luma(1) Both Peak video(1) Both Both Both Pedestal ampl Both Red ampl Both Red phase Both Reference ampl PAL SC/H phase(1) Both Setup NTSC Sync ampl(1) Weighted Frequency Response Both Both noise(1) White b
Measurement Screens (Parameters and Limits) (AVM-510A) EBU/SMPTE Component Measurements (A/B Inputs) Measured Parameter Definition Y ampl Y (luminance) amplitude. B-Y ampl B-Y color difference signal amplitude. R-Y ampl R-Y color difference signal amplitude. Sync ampl Sync amplitude (as encoded back onto signal). Peak video Peak of the luminance signal. H timing Ref Timing of input signal relative to Ext Ref. B-Y to Y Timing of B-Y relative to Y component.
Measurement Screens (Parameters and Limits) (AVM-510A) Definitions of Automeasurement Parameters Measured Parameter Definition 0.5 MHz ref Peak-to-peak amplitude of the 0.5 MHz packet, in IRE or mV. 1.25 through 5.8 MHz Differences between the peak-to-peak amplitude of the frequency packet and the 0.5 MHz packet, in percent. APL Average of active video, aquired in blocks of 10 lines sequenced through the field, in percent.
Measurement Screens (Parameters and Limits) (AVM-510A) Measured Parameter 10-6 Definition K-2 T factor The %K factor if the mathematical representation of the pulse shape that best fits the K factor template equation. Lum linearity Boxes show the amplitude difference between the average of all steps and each step. Any step exceeding 2% shows a + or -. Any step exceeding the limit value turns red. The readout value is the peak-to-peak error (turns red if the limit is exceeded.
Measurement Screens (Parameters and Limits) (AVM-510A) Parameters: Range, Accuracy, and Limits Parameter Range Absolute Accuracy Relative Accuracy Notes Min Setting Max Setting Default Min Default Max 0.5 MHz reference 40 to 120 IRE ± 1 IRE ± 1 IRE 60 IRE nominal 0.5 IRE 100.1 IRE 0.5 IRE 100 IRE 0.5 through 5.
Measurement Screens (Parameters and Limits) (AVM-510A) Parameter Range Absolute Accuracy Relative Accuracy Notes Min Setting Max Setting Default Min Default Max Red ampl Nominal ± 50% Nominal ± 50% ± 1 IRE ± 7 mV ± 1/2 IRE ± 3 mV 88.2 IRE nominal 662 mV nominal 44 IRE 533 mV 148.9 IRE 985 mV 78.0 IRE 582 mV 98 IRE 936 mV Red phase Nominal ± 90° ± 1° ± 1/2° 103.4° nominal 112.9° 112.9° 98.4° 108.
11 11.
AVM-510A Option T (For Transmission Monitoring) The AVM-510A-T senses the measurement signal on the selected line as indicated by the FIELD and LINE designation at the top of the transmitter set. If the correct signal (FCC COMPOSITE or NTS7 COMPOSITE for NTSC or PAL CCIR 17/330 or NATIONAL ITS) is not present, the zero carrier will also not be detected.
AVM-510A Option T (For Transmission Monitoring) Measurement (Composite NTSC, one each for Input A and Input B) SELECT SET SYSTEM VIDEO MEASURE LINE, 1/3:10 to 2/4:262 SIGNAL ID, AUTO SIG ID/LIVE VIDEO/75% COLOR BARS/ 100% COLOR BARS/75% SMPTE BARS/FCC MULTIBURST/ FCC COMPOSITE/NTC7 COMPOSITE/MULTIBURST SYS VIDEO LIMITS SYNC AMPL BURST AMPL PEAK VIDEO PEAK LUMA APL NOISE H BLANKING H REF TIMING BURST REF PHA SC/H PHASE H BLANKING LEVEL COLOR SETUP MEASURE LINE, 1/3:10 to 2/4:262 SIGNAL ID, AUTO SIG ID/LIVE
AVM-510A Option T (For Transmission Monitoring) Measure Signal ID The correct signal must be selected so that the zero carrier and the measured signal will be detected properly. 1. Press the Mesmnt button. 2. Select SELECT SET and then press the Entr button. 3. Select TRANSMITTER and then press the Entr button. 4. Select SIGNAL ID and then turn the knob to AUTO SIGNAL ID. If the signal is known, it can be set to FCC COMPOSITE, NTC7 COMPOSITE, CCIR 17/330 or NATIONAL ITS.
AVM-510A Option T (For Transmission Monitoring) Figure 11-3. Differential Gain Figure 11-4.
AVM-510A Option T (For Transmission Monitoring) To select the differential phase function: 1. Set the center of the lowest point on the waveform to zero. 2. View the center of the peak and calculate the difference in IRE/%. 3. Calculate the differential phase as listed in the following table or use 0.573°/IRE if the reading is under 20 IRE. The graticule in the IRE correlation-to-phase is 0.573° per IRE between 0 and 20 IRE. IRE/% Degrees 2 1.15 4 2.29 7.5 4.3 10 5.7 20 11.5 30 17.5 40 23.
AVM-510A Option T (For Transmission Monitoring) Connections The I (in-phase video) signal from the demodulator is connected to the B CPNT input Y signal BNC connector (see Figure 11-5). The other BNC connector of the loop through is connected with a short cable to the B CPNT input R-Y and terminated at the other connector with a 75 Ω terminator. Note that the I signal is a low pass video signal and that some demodulators only have a video output that can be used with a low pass filter.
AVM-510A Option T (For Transmission Monitoring) Display Setup As stated previously, ICPM modes can be accessed via factory default memories 9 and 10. The following procedures lets you set the display to the correct calibration if the I and Q signals are offset in gain or level (see Figures 11-6 and 11-7). 1. Press the Blue button and then the Mesmnt button to recall memory 9 with ICPM at 2°/division or to recall memory 10 with ICPM at 6°/division. 2.
AVM-510A Option T (For Transmission Monitoring) Figure 11-7. ICPM Display (6°°/Division) Memories Memories have been factory set for applications with the most likely mode of operation. It might be necessary to change some settings in those memories. For example, memory 9 with ICPM 2°/division set to full field display. This needs to be changed to only view a selected line that has the staircase needed for online transmitter monitoring. To do so: 1.
AVM-510A Option T (For Transmission Monitoring) Memory Table AVM-510A-T Memory Locations Memory Input/Format Description Other Items 1 A Composite Waveform, two line display Nonlinear measurement set, line 17 field 1 2 A Composite Picture with save title marks System video measurement set, line 35 field 1 3 A Composite Waveform and vector overlayed 4 A Composite Vector display 5 A Composite Automeasure, transmitter measurement set line 17 field 1 6 A Composite Split screen of picture
12 12. AVM-510A-C (For Component Measurements) Introduction (AVM-510A-C) The AVM-510A-C includes a component measurement set and a digital measurement set when an SDM-550 is interconnected. Component Measurement Sets (AVM-510A-C) Setting limits for component signals SMPTE/EBU, Beta, MII, and RGB can only be done when Input B: is active and is configured for a component signal. To change the Input B: selection to component: 1. Press the Config button. 2. Select INPUT and then press the Entr button. 3.
AVM-510A-C (For Component Measurements) • 75% COLOR BARS forces the AVM-510A-C to recognize the signal as 75% color bars (see Figure 12-1). • 100% COLOR BARS the AVM-510A-C assumes the signal is 100% color bars. 6. Press the down Arrow button to highlight xxx LIMITS, where xxx is the component system currently configured for Input B:. 7. Press the Entr button to continue. Figure 12-1. Setting Limits for Component Signals 8.
AVM-510A-C (For Component Measurements) Serial Digital Measurement Set (AVM-510A-C and SDM-550) NOTE: The digital measurement screen is only active when the AVM-510A-C is connected to an SDM-550. The serial digital measuement set screen provides the following information, including numerical data not present in the picture monitor display (see Figure 12-2). • SIG STRENGTH • JITTER • EDH RATE • PEAK VIDEO To access the serial digital measurement set screen: 1. Press the Mesmnt button. 2.
AVM-510A-C (For Component Measurements) Notes 12-4
A A. Specifications Introduction This appendix lists the the MM-410 Waveform/Vector Rasterizing Monitor (MM-410) and the AVM510A Series Automated Video Monitors (AVM-510A) electrical and mechanical specifications. Electrical Video Standards • NTSC and PAL Composite plus S-video • EBU, SMPTE, Beta, and MII Component (525/625) Inputs NOTE: All MM-410/AVM-510A input connectors are located at the rear panel unless otherwise noted.
Specifications Outputs NOTE: All MM-410/AVM-510A connectors are located at the rear panel unless otherwise noted.
Specifications Measurement Parameters (AVM-510A) AUDIO LEVEL (4 CHANNELS REPORT) COMPONENT COLOR BAR MEAS. (AVM-510A-C) LUMINANCE LINEARITY (PAL) AVERAGE PICTURE LEVEL (APL) COMPOSITE COLOR BAR MEAS.
Specifications Waveform Description Specification Vertical gain accuracy +/– 1% with 1.0 V input Variable gain range >-35% to 15% Vertical Mag x 2.5 typical Frequency response: Flat response Low pass filter response +/– 1% of dc at 50 kHz, < - 40 dB at fsc. Band pass filter response +/– 1% of dc at fsc. Attenuation of > 25 dB at 2 x fsc. 1.5 MHz bandwidth Transient response: Pulse to bar ratio 0.99:1 to 1.
Specifications Vector Description Specification Display accuracy: Amplitude error +/– 2% Phase error +/– 1.25º Chroma bandwidth 700 kHz +/– 100 kHz Variable gain range - 40 to +/– 20% Variable phase 360º Vector MAG x 2.
Specifications Logbook Software Specifications (AVM-510A) • Supported: interconnect RS-232, modem • Baud rate: 9.
B B. Connector Pinouts Introduction This appendix identifies and describes the connector pinouts at the rear panel of the MM-410 Waveform/Vector Rasterizing Monitor (MM-410) and theAVM-510A Series AutomatedVideo Monitors (AVM-510A). NOTE: Connector pinouts are viewed from the outside of the MM-410/AVM-510A rear panel and/or viewed from the wire side of the furnished connectors.
Connector Pinouts RS-232 Connector (Modem/Printing) DCE Pin Number Description 1 Open 2 Transmitted data (TXD) 3 Received data (RXD) 4 Data terminal ready (DTR) 5 Signal ground 6 Data set ready (DSR) 7 Request to send (RTS) 8 Clear to send (CTS) 9 Open 5 1 1 9 6 Figure B-2. RS-232 Pin Locations (Modem/Printing) RS-232 Connector (Computer) DCE Pin Number Description 1 DCD 2 TXD 3 RXD 4 DIR 5 Ground 6 DSR 7 RTS 8 CTS 9 RI 5 1 1 9 6 Figure B-3.
Connector Pinouts Analog Audio Input Connector (AVM-510A) Pin Number Description 1 Analog input 1+ (balanced) 2 Analog input 2+ (balanced) 3 Analog input 3+ (balanced) 4 Analog input 4+ (balanced) 5 Ground 6 Analog input 1- (balanced) 7 Analog input 2- (balanced) 8 Analog input 3- (balanced) 9 Analog input 4- (balanced) Analog inputs 1 through 4 are two-wire inputs. 5 1 1 9 6 Figure B-4.
Connector Pinouts Notes B-4
C C. Memory Locations Introduction This appendix lists the factory set memory locations for the MM-410 Waveform/Vector Rasterizing Monitor (MM-410) and the AVM-510A Series Automated Video Monitors (AVM-510A). To make a change, follow the procedure described in “Saving Settings.” Saving Settings in Memory After you configure the MM-410/AVM-510A, you need to save it in memory.
Memory Locations When a configuration is saved in a memory location, all MM-410/AVM-510A functions are stored in that memory. This includes items that are not displayed. For example, a waveform and measurement setup when a picture is displayed. For best results, recall a memory location, make a modification, and then store the new settings in another memory location. To recall any stored memory at any time, press the Blue (2nd) button and then one of the Memory Recall buttons (for memories 1 through 10).
D D. Calibration Introduction This appendix explains how to access the CALIBRATION menu options for the MM-410 Waveform/Vector Rasterizing Monitor (MM-410) and the AVM-510A Series Automated Video Monitors (AVM-510A). CAUTION: Calibration should only be performed by qualified field service personnel. NOTE: Calibration numbers specific to the MM-410/AVM-510A can be found on an insert supplied with this manual. To access the CALIBRATION menu options: 1. Turn the MM-410/AVM-510A on. 2.
Calibration Notes D-2
E E. Command Codes (AVM-510A) Introduction This appendix describes the RS-232 protocol between the AVM-510A Series Automated Video Monitors (AVM-510A) and a Personal Computer (PC). Guidelines • Letters with ‘ ’ (single quotes) are literal and case sensitive. For example, an ‘a’ is the lower case ASCII letter a, with a decimal value of 97. In general, the case of the letters is significant and must be followed. • means a carriage return. This is ASCII decimal code 13 or 0xD hex.
Command Codes (AVM-510A) PC to AVM-510A (Data Sent) This section describes the data sent from a PC to the AVM-510A. Note that these strings are requests for data or control functions that affect the AVM-510A. Request Functions STATUS REQUEST ‘s’ Refer to “Control Functions” for results. BIT IMAGE SIZE ‘r’ N Set raster image size N = 1 (full size image) N = 0 (¼ size image) Example: r1 causes the AVM-510A to select a ¼ size image.
Command Codes (AVM-510A) BUTTON PRESS ‘B’N. .N N. .N = [0. .63] This is the number of the button as follows: Button Name N. .N BLUE MES_MNT CONFIG UP_ARROW PHASE MENU_CLR PRV_MNU ENTER DOWN_ARROW VAR_GAIN LINE_SEL FIELD_SEL SPLIT_SCRN FILTER DISPLAY HMAG SWEEP VMAG WFM_POS INPUT AUTO_MES PICTURE VECTOR WFM EXT_REF 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 33 Example: B23 is a click on the WFM button.
Command Codes (AVM-510A) KNOB TURN ‘K’N. .N N. .N = [0. .63] The bits of this number are as follows: BIT0 V0 – V0 & V1 are velocity bits BIT1 V1 – 00 = slowest; 11 = fastest BIT2 (not used, leave at 0) BIT3 N0 – N0 & N1 are the knob selector bits BIT4 N1 – 00 = VPOS; 01 = HPOS 10 = MENU/VGAIN/LINESEL PHASE 11 = (not used) BIT5 Set for clockwise rotation; clr for counter (anti) clockwise rotation Example: K43 is a fast clockwise rotation of the H Pos knob. H POS SET ‘H’N. .N N. .N = [0. .
Command Codes (AVM-510A) MEASURE LINE SET ‘N’S. .S TIMECODE LINE SET ‘T’XN. .N INTENSITY SET ‘I’N. .N Also see MEASURE LINE REQUEST The string has the form: “1/3:123” “1/3” an odd field “123” line 123 Also see TIMECODE LINE REQUEST N. .N = [9. .25] Also see INTENSITY REQUEST N. .N = [1. .127] 1 is the minimum PERSISTENCE SET ‘D’N. .N Also see PERSISTENCE REQUEST N. .N = [128. .255] 128 = minimum persistence 255 = infinite persistence VECTOR PHASE SET ‘P’N. .N N. .N = [0. .
Command Codes (AVM-510A) AVM-510A to PC This section describes the data sent from the AVM-510A to a PC. Note that this data might be requested by the PC or it might arrive at the PC unsolicited. STATUS INFORMATION E-6 ‘S’H0H1H2H3H4H5 H6H7H8H9H10 Note: A ‘b’ in front of a bit indicates that it is a 3D button. A ‘l’ in front of a bit indicates that it is an LED status. An ‘m’ in front of a bit indicates that it is a Menu item status.
Command Codes (AVM-510A) H POS NOTIFY H8 m bit0 color frame display on m bit1 VITC time code on m bit2 graticule off m bit3 PAL graticule in percent H9 m bit0 LED reduced size WFM m bit1 LED reduced size VEC m bit2 system in 625 line mode m bit3 measurement set 2 active ‘H’N . .N N. .N = [0. .999] Response to H POS REQUEST. This is a normalized position, that is, a value of 500 would put the slider in the middle of the range. 0 is all the way to the left; 999 is all the way to the right.
Command Codes (AVM-510A) TIMECODE LINE NOTIFY ‘T’S. .S INTENSITY NOTIFY ‘I’N. .N Response to TIMECODE LINE REQUEST. Response to INTENSITY REQUEST. N. .N = [1. .127] 1 = minimum intensity PERSISTENCE NOTIFY ‘D’N. .N Response to PERSISTENCE REQUEST. N. .N = [127. .255] 127 = minimum persistence 255 = infinite persistence PC to AVM-510A (Commands) This section describes the commands sent from a PC to the AVM-510A.
Command Codes (AVM-510A) VCR LOCK SET ‘Y’N N = 0 for crystal lock (narrow) N = 1 for VCR lock (wide) SC/H PHASE DISPLAY SET ‘Z’N N = 0 for off N = 1 for on COLOR FRAME SET ‘F’N N = 0 for off N = 1 for on VITC TIME CODE SET ‘J’N N = 0 for off N = 1 for on GRATICULE SET ‘A’N N = 0 for off N = 1 for on N = 2 for added marks off N = 3 for added marks on PAL GRAT PERCENT SET ‘6’N N = 0 for volts N = 1 for percent REDUCED SIZE SET ‘U’N N = 0 for waveform only N = 1
Command Codes (AVM-510A) PC CONTROL E-10 REPORTING ‘^R’ Sets RS232 and Modem interface to send reports from the REPORT function. CONTROL/ LOGBOOK ‘^C’ Sets RS232 and Modem interface to control the AVM-510A and transfer Waveforms.
F F. Glossary of Terms 1H Display of one horizontal line of video. 2H Display of two horizontal lines of video. Alternate A method of displaying separate signals or separate parts of a component signal in which the parts are super-imposed on top of each other. Blanking The level of picture when there is no picture information at all. The 0 VDC level. In NTSC, blanking is below the black level. In PAL, blanking and black are the same level. BNC The standard coaxial video connector.
Glossary of Terms Chrominance The part of the video signal with the color information. In component and s-video signals, the chrominance is carried in a separate channel. In composite, the chrominance is on a subcarrier in the same channel. Clamping Removing slow changes in DC level by all video levels to the blanking level. Color Burst The brief color subcarrier that immediately follows the trailing edge of sync. Color Frame A sequence of two (in NTSC) or four (in PAL) frames of video.
Glossary of Terms Graticule The reference marks in a waveform display showing horizontal lines for measuring amplitude. One line, the blanking line, has short vertical marks for measuring time. Also, the circular graticule for vector displays, sometimes called a “rose,” that is marked for degree measurements. Internal Reference Using the viewed signal's own sync as the MM-410/AVM-510A’s timing reference. ITS Insertion Test Signal. A test signal in the vertical interval. Same as VITS.
Glossary of Terms S-VHS or Hi8 Video tape formats with separated Y and C channels. S-Video A means of a video connection for separate Y and C channels. Safe Action Area The area within the picture where any action is sure to be seen on any television monitor or receiver. This is usually defined as all the central part of the picture except for anything closer than 5% to the edge of the scan.
Index A about reports, 9-18 added marks, 7-4 alarm outputs special considerations, 9-19 applications AVM-510A option T, 11-10 audio reference levels, 8-2 audio limit reporting, 8-4 audio measurements, 10-4 audio meter, 8-3 audio options, composite NTSC, menu options audio meter, 6-15 audio report, 6-15 ref level dB, 6-15 auxiliaries connecting, 4-2 AVM-510 controlling with a computer, 6-21 features, 3-2 front view, 2-3, 3-3 getting information to and from, 6-22 indicators, switches, and LEDs, 2-3, 3-3 prin
Index mode buttons, 2-4, 3-4 operating, 6-1 power switch, 2-4, 3-4 rack installation, 4-1 AVM-510/MM-410 operation front panel, 5-1 AVM-510A option T applications, 11-10 differential gain and phase, 11-4 ICPM, 11-6 transmitter measurement set, 11-1 AVM-510A option T applications differential gain/phase, 11-10 ICPM, 11-10 AVM-510A-C options component measurement sets, 12-1 serial digital measurement set, 12-3 B blue (2nd button) commonly used functions, 5-12 blue button options auto meas, 5-12 clr mnu (AVM-
Index D G DC control/AVM-510 alarm connecting, 4-3 differential gain and phase AVM-510A option T, 11-4 differential gain/phase AVM-510A option T applications, 11-10 display button front panel, 2-4, 3-4 display options color, 6-3 graticule, 6-4 intensity, 6-3 line select mode, 6-4 persistence, 6-3 safe area, 6-4 split screen, 6-4 wfm limit line, 6-4 display setup ICPM, 11-7 displaying a bowtie signal, 7-14 displaying component signals, 7-11 displaying SC/H phase and time codes, 6-18 displaying waveforms,
Index L labels creating, 6-18 LEDs signal input, 2-4, 2-5, 3-4, 3-5 levels audio reference, 8-2 limit lines, 7-4 limit reporting audio, 8-4 limits setting automeasurement, 9-7 live video automeasurement with, 9-12 locking vector to external sync, 7-18 locking a signal, 7-22 M main menus, 5-8 master alarm 20 second delay, 6-21 turn on/off, 6-20 measure line transmitter measurement set (AVM-510A-T), 11-2 measure signal ID transmitter measurement set (AVM-510A-T), 11-4 measurement sets test signal, 9-3 memor
Index lock, 6-5 serial ports (AVM-510), 6-5 set internal clock, 6-7 P PAL graticule switching, 7-6 parameters automeasurement definitions, 10-5 parameters and limits measurement screens, 10-1 picture changing to external sync, 7-27 comparing two signals, 7-28 viewing, 7-23 power switch front panel, 2-4, 3-4 product overview AVM-510, 3-1 MM-410, 2-1 R readouts color frame, 6-18 reference levels audio, 8-2 report, composite NTSC, menu options error reporting, 6-14 report all errors clear, 6-15 report sustai
Index noise, A-5 operating conditions, A-6 picture monitor output, A-2 power requirements, A-6 safety, A-6 time readout, A-2, A-3 vector, A-5 video inputs, A-1 video standards, A-1 WFM, A-4 sweep rate setting, 7-2 switching PAL graticule, 7-6 T test signal measurement sets, 9-3 time readout, composite NTSC, menu options time, 6-10 time source, 6-10 VITC line, 6-10 transmitter measurement set (AVM-510A-T) AVM-510A option T, 11-1 measure line, 11-2 measure signal ID, 11-4 zero carrier, 11-1 U using filters,