User Manual
20
3. Check the pressure gauge on the regulator. If you have 25-psi or less you will need to replace the gas canister.
4. Place the test Hood from the regulator over the CO sensor.
5. Push Next 3 times to get to the CAL menu then push Enter. The display will flash back and forth between GAS and 200.
6. Start applying gas to the CO sensor.
Note: The sensor will look for the gas for 45 seconds. If no gas is applied or detected in that time, the display will return to CAL.
7. When the sensor detects the gas, the display will flash back and forth between the CO concentration and SPn, then the
calibration will progress and the display will show the gas level for a maximum of 165 seconds.
8. When the calibration is successful, the display will flash back and forth between CO concentration and PAS, then the display
will show the calibration gas level and the calibration is done.
9. If the calibration fails, the display will flash back and forth between the CO concentration and FAL (fail). If this occurs, check the
pressure gauge on the regulator. If the pressure is less than 25 psi the flow of gas may not be adequate to properly calibrate
the unit. If there is proper pressure in the cylinder repeat steps 4 through 6. If the unit fails to calibrate twice contact Technical
Assistance: 1-877-367-7891.
10. Once the calibration has passed, remove gas and disassemble the cylinder and regulator.
11. Re-assemble the CM-12 (make sure the LED is aligned with the front case hole). You are done.
12. See Calibration Flowchart on the inside of the housing.
MACURCO FIXED GAS DETECTION PRODUCTS LIMITED WARRANTY
Macurco warrants the CM-12 gas detector will be free from defective materials and workmanship for a period of two (2) years from date
of manufacture (indicated on the inside cover of the CM-12), provided it is maintained and used in accordance with Macurco
instructions and/or recommendations. If any component becomes defective during the warranty period, it will be replaced or repaired
free of charge, if the unit is returned in accordance with the instructions below. This warranty does not apply to units that have been
altered or had repair attempted, or that have been subjected to abuse, accidental or otherwise. The above warranty is in lieu of all other
express warranties, obligations or liabilities. THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR PARTICULAR
PURPOSE ARE LIMITED TO A PERIOD OF TWO (2) YEARS FROM THE PURCHASE DATE. Macurco shall not be liable for any
incidental or consequential damages for breach of this or any other warranty, express or implied, arising out of or related to the use of
said gas detector. Manufacturer or its agent’s liability shall be limited to replacement or repair as set forth above. Buyer’s sole and
exclusive remedies are return of the goods and repayment of the price, or repair and replacement of non-conforming goods or parts.
Manufactured by Aerionics, Inc.
Sioux Falls, SD
Email: info@aerionicsinc.com
Phone: 1-877-367-7891
Rev 08.19.2014
© Aerionics 2014. All rights reserved.
Macurco is a trademark of Aerionics, Inc.