Technical data
72
131.
DNA Analyzer & Sequencer
(APPLIED BIOSYSTEMS/LICOR/EQUIVALENT WORLDWIDE RECOGNIZED
MAKE
SYSTEM)
• Fluorescence-based four capillary (upgradable to 16 capillary) Genetic Analyzer / DNA analysis system to support
a variety of molecular applications including sequencing and DNA fragment (AFLP; SSR; SSCP ; SNP) analysis
with resolution of up to 1bp;
• Fully automated sample loading to data analysis;
• Electrophoresis voltage should range from 100 V up to 15 kV with operating temperature range of the capillaries
suitable for sequencing and DNA fragment analysis;
• Argon-ion laser excitation based CCD detection;
• System capable of using /96-well plate kits;
• Instrument based on liquid flowable polymers as the separation matrix with automatic replenishment of the
capillary;
• Multiple lengths capillaries for multiple applications;
• Up to 5 dye detection system; upgradeable to incorporate future dye sets and chemistries without any hardware
changes;
• Supplied with necessary consumables for 2000rxn DNA fragment analysis and 1000rxn of sequencing (polymer;
buffer; sample tubes; septa; size standards, deionized formaimide;
• Accessory spare parts viz. electrodes, valves; syringes etc. should be supplied in duplicate;
• The system should be supplied with ddNTP based sequencing kits for long run; DNA Fragment Analysis
Reagents: (Heat Start Taq DNA polymerase, AFLP preselective & selective amplification kits, SNP kits)
• Supplied with a computer workstation for instrument control and data analysis loaded with necessary software for
sample ID import, instrument control and data collection; Computer with minimum 80 GB HDD,512 MB RAM,
Built in DVD-RW/CD-RW combo drive and 19inch Flat panel display and high resolution color deskjet printer
should be provided.
• Sequencing Software for Basecalling and Gene fragment Analysis software to perform fragment size detection,
linkage & mapping software, heterozygosity detection software; SNP detection software.
• Softwares for fragment sizing and alignment of analyzed samples together;
• Power Requirements: Voltage 200-220V or 230-240V +/- 10%, 50/60HZ +/- 10 % Current: Maximum 15 amps.
• It should be supplied with electric backup system based on lead batteries capable of handling the system in power
failure for atlease 6 hours.
• The company supplying the instrument should have proven capability demonstrated in the past in after sales
service and complete support as for as the
• Application systems of the machine.
• The supplier should have the facility for comprehensive training on the operation of the instrument, software and
Applications Support in India.