Technical data

37
55. BOD Incubator Automatic Steelmet make:
BOD Incubator Automatic
Works Space: (Interior dimensions)- 18” (D)x24”(W)x36(H), Temp. Range: +5
o
Cto60
o
C (heating and colling
both), Accuracy±0.5
o
C, Temp. Controller: with PT-100 sensor with digital display
Construction-
Interior: 18 swg Stainless Steel 304 sheet
Exterior: 18 swg CRCA(MS) sheet-powder coated Viewing toughened double glass window to avoid temperature
losses. Temperature shoot ups warned through audio alarm buzzer. Chamber lights ON-on oerating the door
Utilities: S.S. adjustable trays will be provided along with the incubator.
Regrigeration system: ‘Kirloskar’ make hermatically scaled compressor, ‘Danfoss’ make solenoid valves, cooling coil
pressure gauge etc. Time delay to safefuard compressor.
OR
OSWORLD ISO 9001 COMPANY MODEL-JIRC 10, METALAB / YORKO / Deep Instruments
Electronically controlled double walled in construction exterior made of mild steel with powder coated finish, interio
made of stainless, glass wool insulation with magnetic gasket and lock, consist digital controller with 0.1
0
C with
sensory, provided with air circulation for temperature controller with chamber,
For GMP model/ Kumar make/ /steel met make
These Microprocessor controlled cooling incubators are designed especially for close control of incubation / storage
temperatures.
These chambers are with soft touch keyboard for fast and accurate setting of various parameters. Constructed of double
walls, the exterior is made of sheet steel while the inner chamber is made of stainless steel. GMP model with inner
chamber of SS 316 and outer body of SS 304 stainless stee.
Salient Features :
* User friendly Microprocessor Design / Operation with precise temperature control
* Safety thermostat to prevent over heating.
* Audio visual alarm warning of Temperature variation.
* Censorings interface for keeping hard copy of Temp. Time and Date.
* Non volatile memory for data storage upto 620 records.
* Adjustable print interval.
* High voltage safety cutoff for unit protection.
* Flexi glass inner door to inspect specimenInterface for temperature control through personal computer and for
multiple
networking of chambers.
Chamber illumination with fluorescent or ultra violet lights.
Cyclic timer for regulating illumination conditions.
6 Cubic Ft. Temp. 5°c to 60°c inner chamber of S.S. with Centronic printer interface 10 Cubic Ft. Temp. 5°c to 60°c
inner chamber of S.S. with Centronic printer interface
12 Cubic Ft. Temp. 5øc to 60øc inner chamber of S.S. with Centronic printer interface