47. Bioanalyzer:
Agilent Technology make : 2100 Bioanalyzer
(CAT No. G 2940CAs)
2100 bioanalyzer desktop system – Includes 2100 bioanalyzer instrument, HP Compaq PC, 17” Monitor, HP DeskJet
Colour Printer, 2100 expert software, vertex mixer with chip adapter, and chip priming station, without cartridge and
license
2100 Bioanalyzer Electrophoresis Set (CAT No.G2947CA)
Includes Electrode cartridge, License key for electrophoresis assays, test chip bundle and start-up service.
OR
Make: Shimadzu Biotech,Japan, Model: MultiNA
Simultaneous up to 4 MicroChip loading for analysis with a cycle time of 75 sec & < 25 min analysis for 12 samples
1) Reusable Micro chips for more than 1500 analysis
2) Compatibility with wide range of sample tubes / plates for automated sample loading.
3) Capability to analyze very low sample volumes from 2 to 30 ul
4) Minimum of 100 analysis per analysis schedule
5) Minimum of 96 number of samples loading
6) Detection method: Fluorescence, filter spectroscopy (cut-on wavelength 525 nm)
7) Excitation light source: Blue LED (center wavelength 470 nm )
8) Detector: PMT (Photomultiplier tube)
9) Temperature during electrophoresis: 37°C
10) Electrophoresis power supply: Maximum rated voltage 1.5kV, maximum current 250 µA
11) Separation buffer filling: Offline pneumatic system (air cylinder)
12) Separation Kit for DNA/RNA analysis :- 500 to 2500 bp DNA analysis kit with sizing resolution of < 10% &
sizing accuracy of < 15% & RNA kit for RNA analysis
Software for one point instrument control , operation & data analysis
a) Analysis:Analysis schedule creation, real time control, automated analysis preprocessing, automated analysis post-
processing, automatic error processing
b) Display monitor, log: Real time display, analysis log maintenance, usage log for consumable items and
maintenance parts
c) Maintenance service: Analysis performance check, checks following replacement of consumable parts
d) Data formats: Proprietary data format, sample information import using CSV / EXCEL format
e) Data analysis: Sizing, quantitation, manual editing and reanalysis of analysis results
f)
Data display: Gel image / electropherogram list viewing / detailed display, peak table / RNA report display
Output: Analysis information, gel image, electropherogram, analysis results can be printed in various layouts or
exported in CSV format
Utilities: Data search, display of analysis performance check results