Technical data

22
Removal, Fully Automatic Water level Monitoring, 20 Predefinable user storage programmes with auto sequence, 24
keys, 32 characters, 2 line back-lit Alpha Numberic Display, 20 different users Programmable code on a single system,
Distillation Capacity: 15-20 Samples/hr, Nitrogen Recovery > 99.5%, Stainless steel non corrosive steam Generator,
RS 232 interface for connectivity to Printer & PC, Software for documentation of results with builtin formulae
for 9 parameters, Builtin safety features like over temperature protection sensor, over pressure control valve, etc.
Electrical Requirement – 220V/50Hz.
IV. Automatic Digital Burette Titration System
Compact Titration System, End point of Titration controlled by 2 non slip hand wheels from continuous to dropping
speed, Accurate 2nd decimal digital display at equivalence point, Volume ml - 50ml, subdivision ml - 0.01, Accuracy -
+
0.2 ml, Piston type with safety prime valve, telescoping filing tube, Conformity certified with electronic volume
display, Battery pack for 60,000 titrations
26. Automatic Nitrogen Analyser Programmable Automatic Distillation Unit with built-in AOAC approved colorimetric Titration with built in program
for 30 Pre-defined official methods & 10 customizable method, CE Certified from an ISO 9001 company based on
KJELDAHL TECHNIQUE consist of :
a) Digestion Unit for 20 tubes: Programmable Digestor System temperature
control by microprocessor, LCD display, RS 232 connectivity.
b) Distillation Unit: Programmable Automatic Distillation Unit with integrated
Colorimetric Titration System (30 standard pre-defined programs & 10 customizable
programs – all parameters concerning distillation & titration are programmable) having
automatic alkali addition, automatic water dilution, automatic boric acid addition and
automatic distillation residue removal, steam flow regulation 10 to 100%. RS-232
connectivity to a printer or PC for data storage and should have alkali resistant
technopolymer splash head for longer life. Should have maintenance free steam generator
with patented titanium condenser for efficient heat exchange. Should include safety
features like safety door, tube in place safety and water saving system. The method adopted
by the instrument should be in accordance with AOAC, EPA, DIN & ISO. Complete with 1
Digestion test tube 300 ml, 1 collecting flask 250 ml, 1 pincer for test tube, inlet tube, set of
tubes & keyboard.
Distillation Time : Starting from 4 minutes, Measuring Range : 0.1
tto 200 mg Nitrogen
Nitrogen Reproducibility :
1%, Recovery : > 99.5%
Alkali Addition : 0-100 ml, Dilute water addition : 0-200 ml
Boric Acid addition : 0-100 ml, Storage Capacity : Upto 3500 values
Maximum Burette Standard Value : 25 ml