Direct Drive Mechanism on arms, less maintenance and Accurate Position for Tips locations.
The system should have the flexibility to have the password protection in three levels user, admin and integrator.
The pipetting tips should have the Liquid detection features as standard for both Washable tips & Disposable tips. Disposable tip
sensing functions for tip pick up, during operation up to the end of the process.
Carrier to reagent : The following reagent carrier with required plastic consumable troughs to be quoted separately.
• 200 ml,100ml,25ml trough to place different type of reagent required for DNA/Nucleic acid isolation and ELISA
processing.
• Carrier for test tubes and eppendorf tubes.
• Disposable tips in the optional.
• Please quote the trough and carrier required for ELISA processing set up.
Also the system should have the facility to increase the throughput by upgrading the unit, such as hotels, shelf and with other
storage vessels in the same platform to store microplate, filter plate etc.,
The system should have the microplate Reader with tunable wavelength in 1nm increment from 400-700nm as well as should have
8-channel microplate washer(unpressurized system) with the residual volume of less than 2ul for fully automated ELISA
processing.
It should be integrated with incubator with 6 position capacity for ELISA plate incubation.
The system should be supplied with software for full access to application development environment as well as required Computer
& Printer.
Similar liquid handling system should be supplied installed minimum of 5 nos in India and performance certificated and
communication details should be provided.