Technical data

128
Contd….
C. Sample Introduction System
1. Nebulizer: Concentric type glass nebulizer,and inert HF-resistant nebulizer
2. Spray Chamber: quartz cyclonic spray chamber, inert HF-resistant spray chamber
3. Torch: Low flow, low power, demountable,
4. Peristaltic Pump: Computer-controlled, dual channel 12-roller pump, at variable speed 0-60 rpm.
5-Torch-should be made of quartz single piece and should be vertical(Radial)
D. Detector
PMT to cover the spectral range: 160 nm to 800 nm. Wavelength Calibration Characterize the full wavelength range
E. Instrument Control
IBM compatible PC based for instrument control, data acquisition, data processing and report generation.
Computer system:
Latest configuration Pentium IV
21" TFT colour, non-interlaced rnonitor
Colour HP DeskJet printer,
8. Windows XP professional operating system
1. Type: Windows based, multi-tasking powerful and flexible software providing full control over all instrument
functions (including pump and accessories like auto-sampler) data collection and processing.
2. Analysis Mode-Automatic, Manual and semiquantitative
3. Wavelength database: Built-in wavelength database
4. Parameter control: Computer control of viewing height, viewing centre, power, plasma, auxiliary and nebulizer gas
flows,speed, integration time, scan window , PMT voltage, Customizable instrument status display.
5. Correction modes: Fixed point or dynamic background correction. Internal standardization. Inter-element
correction.
6. Automatic run options: Automatic plasma shutdown. Automatic saving of all data. Stability check testing
7. Data handling: Report generation, results viewing. Export results to third party software.
F. Exhaust system: Custom designed dedicated exhaust
G. UPS: Fully compatible UPS to be quoted separately in Indian currency indicating the complete technical
specifications and part number.
H. Spare consumable and accessories: Essential spares and consumables required for smooth uninterrupted operation
of the instrument for a period of one years to be quoted separately.
I. Installation, commissioning and Training: Bidder shall provide pre-installation advice after finalization of the
offer.
Successful bidder shall provide comprehensive advanced operation and maintenance training to our
scientists/Engineers
J. Typical detection limits in ppb at 3 sigma: Detection limits should be provided for all elements