Hunter Lab Color
Spectophotometer Color flex
A bench top color measuring system (HUNTERLAB COLORFLEX Make) for the measurement of color of food &
dairy products in various forms (solid, liquid, powders, paste, gel, whole fruit etc. ) with the following Measurement
Principal : Dual beam spectrophotometer
Measuring Geometry : 45/0,
Wavelength Range : 400 – 700 nm
Pitch : 10 nm,
Port diameter : 31.8 mm
Wavelength Resolution : < 3 nm
Photometric Range : 0 – 150 %
Light Source : Pulsed Xenon Lamp,
Lamp Life : > 500,000 flashes,
Measurement Time : < 1 sec
Standard accessories : Calibrated instrument Tiles, Certificate of traceability, Diagnostic tile, black glass
Optional Accessories : Glass sample cup, Sample cup set, EasyMatch QC Software,
227.
Hybridization Chamber
• Horizontal desktop shaking platform with two independent incubation chambers each capable of working as an
independent unit (with 2 different temperatures).
• Hybridization incubation chambers should maintain temperature (range from within 10˚C of ambient to at least
80˚C) with platinum sensor and control accuracy +
0.1˚C.
• Temperature display system with input board for adjusting input parameters during hybridization.
• Chamber capable of handling at least 4 hybridization bottles of variable size ranges
• System should ensure continuous flow of minimum volume of hybridization buffer across the complete surface of
the hybridization membrane.
• Rotation speed should be adjustable as per different requirements.
• Complete set of hybridization bottles (Radiation and leakage proof) of variable ranges (set large-25-35cm long;
medium15cm long and small 10cm long) with diameter (3-5cm) enough for allowing easy insertion and removal
of membranes.
• Metal framed, double glass vacuum doors to protect user from radioactive exposure and system to prevent
accidental spillage.