Technical data
111
206.
Growth Chamber with
Controlled Humidity,
Temperature and Light
Intensity
Percival Scientific, Inc. USA , Model : LT-105
Temp Range: 0-44
0
C±0.5
0
C, Interior Space Volume : 76.4 ft
3
(2.2m
3
)
Total Shelving Floor Area : 16.1 ft
2
(1.5m
2
), Maximum Growing Height: 57 in. (142.2cm)
Exterior Dimensions in inches (cm): Height : 77.2(106.1) ,Width : 117(297.2), Depth : 37.1(94.2)
Light intensity: 980 µmol/m
2
/s, Number of Tiers : 1
Advanced Intellus System (C9), Communication Software(C9+), Advanced Intellus with Touchscreen and Internet
capabilities (C10), CO
2
enrichment package, self contained air-cooled condensing unit,remote air cooled condensing
unit, dry alarm contacts ( S-2), closed loop dimmable lighting ( Q -22), open loop dimmable lighting ( Q -23), extended
temperature ranges.
OR
Sheel Biotech make, Model - 03605 Interior Space Volume : 76.4 ft
3
(2.2m
3
)
Pre Fabricated Cabinet -
Double walled cabinet structure with Dirt repellent Powder Coated GI 0.5mm thick, outer and Stainless Steel (
304 grade ) non magnetic inner surface
60 mm polyurethane ( Puff ) insulation ( Puff Density : 42±2 Kg/cum. ).
Prefabricated panel system will be duly sealed to prevent leakage.
A Flushed door (3’ x 2’) duly insulated will be provided. The door fitted with heavy duty Gravity Hinges
(Imported). with sweep gasket to prevent leakage will be provided.
The door fitted with heavy duty hinges & lock system
See through hatch window of 25cm x 25cm size (Toughened Vacuum Glass Panel fitted with aluminum/ SS
sections) on two sides will also be provided.
Additional Sub Door of Glass 5mm thick with lock & key arrangement will also be provided
Castor arrangement will be provided.
Cooling/Heating System
Range: -5°C to 50°C-70°C ± 0.5 °C without humidity& illumination
Range: 25 °C to 60°C ± 1.5°C without humidity& illumination.
Cooling down time to 10°C – 35 minutes minimum
AC System (Sheel Make):-
“For efficient working of the Growth Room separate cooling chamber will be provided”
Humidity Range : upto 95% +
2%
Through AHU Unit, it creates the simulated natural humidity in Growth Chamber because humid air comes out in cold
form.
Illumination System ( Door/ above or below the each selves) : Photo synthetically Active Radiation
Lamp (P Ac Ra) : MCS (Microprocessor control System) for Growth Chamber”Programmable Day & Nights
separate setting for week program
Microclimatic Temperature Controller {Code No. S 0256}
Microclimatic Humidity Controller { Code No. S-
0274} Photoperiodic Timer Programmable {Code No. S-0264}
Shelves - 4-6 Heavy duty removable Stainless Steel shelves with rounded inner chamber corner for easy cleaning will
be provided inside the Chamber.