Technical data

108
199.
Glassware Washer with
Drying System
Steelco Spa Italy, Model :500C
Programmable laboratory glassware washer & dryer for washing and air drying of different types of glassware. Should
be designed to comply with latest international guideline on decontamination.
Double wall construction and excellent insulation should be provided to reduce heat loss and power consumption.
External cabinet should be made of 304 stainless steel. Inner cabinet, washing arms and tank filters should be of high
quality AISI 316L stainless steel. Construction system with total removal of angles. Self cleaning tank with rounded
edges. Water filling system in three levels enables residue to be captured and not recirculated to saves the pump life.
Three water connections should be available, cold, warm & DI water. There should be two automatic liquid dispenser
and with an option for one more dispenser. HEPA filtered forced air drying system with adjustable time and temperature
setting to ensure completely dry, inside and outside of all glassware. RS 232 port should be provided with printer
connection for monitoring and validating washing process. LED display with five adjustable washing programs. The
control system should have automatic diagonistic checking which constantly monitors and displays current cycle status
and alarms.
Should be supplied with inserts for 30 positions petri disches, upper & lower baskets with 36 cleaning jets each for
cleaning bearkers & flasks, upper mixed basket for centrifuge tubes, vials and test tubes of 121 positions and lower
vertical pipette basket of 120 positions.
200.
Global Positioning System
(GPS)
i) Sportrak map Accuracy to within 3 meters, map features to 2 MB North American Mapping database
ii) Mobile mapper
OR
Garmin Handheld GPS etrex Vista Model with software
201.
Gradient PCR
Silver Heating Block – 6 degree C / sec and cooling 4.5 degree C / sec
Gradient range – up to 24 degree C
Gradient possible at all three stages (denaturation, annealing & elongation) in one program.
Electronic sample protection – Heated lead should fit on sample only after reaching set block temperature.
Electronic hot start – Initial heating ramp up to 12 degree C / sec
Control panel – Windows type display; capable of controlling up to five heating blocks.
PCR License
Upgradeability options - to QPCR, Networking of multiple blocks to one console, Automisation of Heating lid
OR