Instructions

en
NANOCOLOR
®
Ammonium 3
MACHEREY-NAGEL GmbH & Co. KG · Neumann-Neander-Str. 6–8 · 52355 Düren · Germany
www.mn-net.com
DE / International:
Tel.: +49 24 21 969-0
Fax: +49 24 21 969-199
E-mail: info@mn-net.com
CH:
Tel.: +41 62 388 55 00
Fax: +41 62 388 55 05
E-mail: sales-ch@mn-net.com
FR:
Tel.: +33 388 68 22 68
Fax: +33 388 51 76 88
E-mail: sales-fr@mn-net.com
REF: 985003
Overview
The test is suitable for the photometric determination of NH
4
+
.
The test is in accordance with APHA 4500-NH
3
F, ISO 7150-1,
DIN38406-E5 and EPA350.1.
The test is suitable for surface water, groundwater and drinking
water and slightly contaminated wastewater.
Results are highly reproducible in water with low levels of pollutants.
Measuring range:
0.04–2.30mg/L NH
4
-N (method 0031)
0.05–3.00mg/L NH
4
+
/ NH
3
(method 0032 / 0033)
Number of tests: 20
Wavelength for photometric determination: 660/690nm
Shelf life: 12 months
Reaction time: 15 minutes
Storage temperature: 15–25°C
Storage conditions: upright
Method
Photometric determination of a blue indophenol dye based on
a reaction with hypochlorite and salicylate at a pH of 12.6 in the
presence of sodium nitroprusside.
Interferences
Heavy pollution results in errors and requires prior distillation.
The method is suitable for the analysis of seawater after 1+1
dilution.
Turbidities cause higher measurement values.
Reagents and accessories
Contents of reagents set:
20 test tubes R0
1 NANOFIX R2
Required devices:
MACHEREY-NAGEL photometer
Digital piston pipette 1–5 mL (REF916909) with pipette tips
(REF916916)
Tweezers for sampling NANOFIX capsules (REF916114)
Standards
NANOCONTROL Multistandard Sewage outflow 2 (REF925010)
Sampling and preparation
See DIN EN ISO 5667-3-A 21.
Adjust to pH 7–10 with NaOH or HCl prior to analysis.
Adjust highly buered solutions to pH 9–10.
Quality control
The measurement of a blank value and a standard is recommended
before every measuring series as quality control measure.
Quality data:
The following data were determined during production according to
ISO 8466-1 and DIN 38402-A51:
Number of LOTs: 40
Standard deviation of the method: ±0.02mg/LNH
4
-N
Coecient of variation of the process: ±1.60%
Confidence interval: ±0.04 mg/LNH
4
-N
Specified data for procedure:
Sensitivity (absorbance of 0.010 A corresponds to):
0.013mg/LNH
4
-N
Accuracy of a measurement value: ±0.03mg/LNH
4
-N
LOT-specific certificates are available at www.mn-net.com.
Procedure
1. Open test tube
2. Pipette 4 mL of sample into test tube
3. Add 1 NANOFIX R2
4. Seal test tube and shake vigorously
5. Wait 15 min
6. Clean outside of test tube
7. Measure
Notes
When using other photometers, make sure measurements are
possible in test tubes (16 mm OD) and calibrate the method.
Correction value e. g. for colored or turbid samples possible (see
photometer manual).
Information regarding safety can be found on the box' label
and in the safety data sheet. You can download the SDS from
www.mn-net.com/SDS.
08/2020