Hardware manual

Maintenance 4
Agilent 7500 Series ICP-MS Hardware Manual 4-11
Recording the Maintenance Log
When the maintenance logs are recorded, current meter values and the last
tuning report are recorded.
To record the maintenance logs, complete the following steps:
1 Click Record Log in the Maintenance LogBook window.
The Logbook Editor dialog box appears.
Figure 31 Logbook Editor dialog box
2 In the Logbook Editor, if there are maintenance items to be done at this
time, click the check boxes. You can also type comments to record in the
logbook.
3 Click OK to record the maintenance log.
The Maintenance LogBook is created. The tuning parameters and tuning
result values when the tuning report was printed are saved. Meter values
are saved by clicking OK.
NOTE
The maximum number of logs is 256 per day.