Hardware manual
4-4 Agilent 7500 Series ICP-MS Hardware Manual
4Maintenance
* The frequency of maintenance indicated for these parts is a general guideline only. Actual
cleaning regimen may be infrequent, dependant on local lab/workload variables.
Refer to the Agilent 7500 Series Maintenance DVD for instructional videos on how to perform all
maintenance tasks. It is easier to schedule maintenance if you keep a log of the instrument
readings for each analysis. Generating a tuning report daily is also strongly recommended. The
“Maintenance LogBook” feature explained later in this chapter has a similar function.
Periodically
(Maintenance frequency of
these components is highly
dependent on lab conditions and
sample throughput. They should
be checked periodically, at least
on an annual basis, and
appropriate action taken)
Sample Introduction area
parts: such as Spray
chamber, End cap
Clean
Torch Clean Replace when necessary
Electron Multiplier Evaluate Replace when necessary
Plasma gas, auxiliary gas
tubing
Check Replace when necessary
Argon gas filter Replace (2 years after
installation)
Tabl e 9 Maintenance Schedule (continued)
Frequency Component Task/Action Remarks