Hardware manual

Startup, Shutdown and Status 3
Agilent 7500 Series ICP-MS Hardware Manual 3-19
4 In Standby mode
5 In Analysis mode (std) (7500cs/cx)
Tabl e 7
Parameters Limit Conditions
Analyzer pressure (AN) > 2 x 10
-3
Pa Agilent 7500 goes back to Shutdown mode.
Backing pressure
(IF/BK)
> 60 Pa Agilent 7500 goes back to Shutdown mode.
Tabl e 8
Parameters Limit Conditions
Interface pressure
(IF/BK)
> 530 Pa Agilent 7500 goes back to Standby mode.
Analyzer pressure (AN) > 2 x 10
-3
Pa Agilent 7500 goes back to Standby mode.
Backing pressure
(IF/BK)
> 60 Pa Agilent 7500 goes back to Shutdown mode.