Hardware manual
3-16 Agilent 7500 Series ICP-MS Hardware Manual
3 Startup, Shutdown and Status
4 To monitor other components, deselect the current ones by clicking the
appropriate boxes. Then click the boxes next to the additional
components you want to monitor.
Meter boxes for the newly selected components appear on the right side of
the screen.
5 Repeat Step 4 until all necessary components have been checked.
Tabl e 3 Meter Control Panel Values
Meter Typical Ranges Units
Shutdown Mode Standby Mode Analysis Mode
IF/BK Press – 0.3 ~ 5 250 ~ 490 (1 RP )
*1
180 ~ 300 (2 RPs)
Pa
Analyser Press – 3x10
-5
~ 6x10
-4
3x10
-4
~ 2x10
-3 *2
Pa
TMP Revolution – 65 ~ 100 95 ~ 100 %
Water RF/WC/IF 0.0 0.0 1.3 ~ 5.0 L/min
Water Temperature 15 ~ 25 15 ~ 25 15 ~ 45
o
C
Inlet Temp Room Temp. Room Temp. 25 ~ 35
o
C
Exhaust Temp Room Temp. Room Temp. 45 ~ 55
o
C
S/C Temperature (L) Room Temp. Room Temp. 2
o
C
Carrier Gas (BP)
*3
0 0 450 ~ 600 (BN)
*4
200 ~ 300 (CN)
200 ~ 500 (MFN)
200 ~ 300 (MMF)
kPa
OP Gas Press 0 0
*5
kPa
OP Gas Tank Press 0 0
*5
kPa
Ar Gas Tank Press 700 700 700 kPa
Carrier Gas 0 0 0.8 ~ 1.3 L/min
Makeup Gas 0 0 0 ~ 1.0 L/min
Optional Gas 0 0
*5
%
Plasma Gas0015L/min
Aux Gas 000 ~ 1.0L/min
Forward Power 0 0 700 ~ 1600
*6
W
Reflected Power 0 0 < 20 W