Hardware manual
3-4 Agilent 7500 Series ICP-MS Hardware Manual
3 Startup, Shutdown and Status
6 Startup the printer and computer.
Refer to the printer and computer user manuals for more information.
7 Start the MassHunter Workstation (or ChemStation) software.
Refer to Chapter 1, “Using the Agilent 7500 Series ICP- MS MassHunter
Workstation Software” in the Agilent 7500 Series ICP- MS MassHunter
Workstation Operator’s Manual (or “Using the Agilent 7500 Series ICP- MS
ChemStation Software” in the Agilent 7500 Series ICP- MS ChemStation
Operator’s Manual) for information on accessing and exiting the software.
The ICP- MS Top window appears.
Figure 16 ICP-MS Top Window
CAUTION
When the MassHunter Workstation (or ChemStation) software starts running
immediately after turning on the main switch of the ICP-MS instrument, the mainboard
of the instrument sometimes hangs up and becomes unavailable to control the
instrument. If this happens, turn off and then on the mainframe, wait more than 40
seconds, and then turn on MassHunter Workstation (or ChemStation). MassHunter
Workstation (or ChemStation) software should be started more than 40 seconds after
turning on the main switch of the instrument.