Agilent 7500 Series ICP-MS Hardware Manual 2-15
7 Changes or modifications not expressly approved by Agilent
Technologies could void the user’s authority to operate the equipment.
ICES/NMB-001
This ISM device complies with Canadian ICES- 001.
Cet appareil ISM est conforme à la norme NMB- 001 du Canada.
Electrical Power (Voltage, Frequency, Amperage, Phase)
200- 240 VAC, 50/60 Hz, 30 A, single phase
Argon Gas Supply
Minimum Purity: 99.99%
Maximum Flow Rate: 20 L/min
Supply Pressure: 700 kPa
Reaction Gas Supply (Agilent 7500cs, 7500cx)
Gas: Helium, Hydrogen (optional for 7500cx)
Minimum Purity: 99.999%
Maximum Flow Rate: He 10 mL/min, H
2
7mL/min
Typical Pressure: 40 kPa
Cooling Water
Inlet Temperature: 15 - 40 °C
Inlet Pressure: 230 - 400 kPa
Minimum Flow Rate: 5.0 L/min
Exhaust Duct
Extraction Flow Rate: 4 - 6 m
3
/min Agilent 7500a/cx (without optional
duct) 7 - 8 m
3
/min Agilent 7500ce/cs (with optional
duct which is standard on 7500cs)