Hardware manual

Precautions 2
Agilent 7500 Series ICP-MS Hardware Manual 2-13
Moving the Agilent 7500 ICP-MS
Environmental Conditions and Utility Requirements
Temperature and Humidity Ranges
The Agilent 7500 ICP- MS should be operated and stored under the
temperature and humidity ranges described below:
Operation condition:
Temperature: 15 °C - 30 °C (59 °F - 86 °F)
< 2 °C change in 1 hour
Relative humidity: 20% - 80% (non- condensing)
Altitude: Up to 2000 m
Atmosphere: Non Condensing; Non Corrosive
For optimum instrument performance, the temperature variations should be
less than 2 °C in 1 hour.
Storage condition:
Ambient temperature: - 20 - 60 °C
Relative humidity: 15 - 85% (non- condensing)
Installation category based on IEC1010: II (See Note.)
Pollution level based on IEC1010: 2 (See Note.)
CAUTION
Stated below are precautions you should observe when you want to move your Agilent
7500.
Make sure the Agilent 7500 Main Power Breaker (located on the rear side) is turned
OFF before moving the Agilent 7500.
Make sure all the cables between other units are unplugged and the utility tubing
are disconnected before moving equipment (small moves for adjustments etc. can
be done without unplugging and disconnecting).
The Agilent 7500 is heavy (about 175 kg), so if there is a need to lift the Agilent 7500,
it should be lifted by at least by 4 people or with a mechanical lifter.
At time of installation, and upon instrument movement, please check for leaks using
appropriate device.