Allowing the Hot Instrument to Cool
Torch Box Cover
Peristaltic Pump
Rotary Pump
Air Intake and Outlet
WARNING
The torch and interface remain hot after the plasma is turned off. Do not touch the
torch box or interface cones for 10 minutes after turning off the plasma to allow for
cooling.
CAUTION
Attach the torch box cover after maintenance. This cover reduces the emission noise
from the plasma.
CAUTION
Be aware that the clamp of the peristaltic pump opens and closes. You can get your
finger stuck in between the stator and the rotator, or between the stator and the
stopper.
WARNING
Rotary pump oil is flammable. Keep away from fire.
For accidental spillage of rotary pump oil on skin, mouth, eyes; wash immediately
and thoroughly and seek expert medical attention.
CAUTION
Do not cover the holes for air intake and outlet on the Agilent 7500.