Hardware manual

Overview 1
Agilent 7500 Series ICP-MS Hardware Manual 1-21
Support Systems
The Agilent 7500 requires argon, cooling water, electrical power and positive
exhaust ventilation. Refer to the Agilent 7500 Series ICP- MS Site Preparation
Guide for additional information.
The MassHunter Workstation (or ChemStation) software controls the Agilent
7500 and interprets the resulting data.
Following subsystems are water cooled.
Spray Chamber
Work Coil
Interfaces
RF Generator
Optional Equipment
The Agilent 7500 can be adjusted for various types of analyses by utilizing
options in the sample introduction area. For details of the optional equipment
please refer to the
Agilent 7500 Series ICP- MS Customer Maintenance Parts
List
.
NOTE
If an anti-freeze solution is added to the cooling water, do not exceed 10% v/v. It will reduce
the cooling efficiency and may overheat the components of the Agilent 7500.