Hardware manual
1-14 Agilent 7500 Series ICP-MS Hardware Manual
1Overview
Sample Introduction
The Agilent 7500 features an on- board peristaltic pump that controls the flow
of sample solution into and waste (drain) out of the instrument. The
peristaltic pump continuously pumps the sample solution into the sample
introduction system. The solution passes into a nebulizer that disperses the
sample using a stream of argon gas (Figure 9).
Figure 9 Nebulizer and Spray Chamber (Top: Babington type, Bottom: concentric type)
A sample solution mist forms and passes through a double- pass spray
chamber where the larger sample droplets are removed by collision with the
spray chamber wall. To prevent the spray chamber from filling up with liquid,
the peristaltic pump drain channel constantly removes excess sample solution.
The Peltier cooler, a thermoelectric device, cools the spray chamber. This
cooler enables the temperature inside the chamber to be precisely controlled,
which gives the Agilent 7500 a very stable ion signal. In addition, cooling the
sample aerosol also removes some of the water from the sample, reducing the
level of polyatomic oxide species formed, which in turn reduces the
interference on certain analytes.
To Torch
Drain
Spray Chamber
Nebulizer
Carrier Gas
Sample
To Torch
Drain
Spray Chamber
Nebulizer
Carrier Gas
Sample