Hardware manual

A-8 Agilent 7500 Series ICP-MS Hardware Manual
Appendix
The following explains the different points between the standard drain kit and
the drain kit for Organic Solvent.
(2) Bushing: Material is different.
(3) Tubing: Material and ID / OD are different.
(4)(6) Connector: OD to connect the Tubing (3) / (7) is different.
(7) Tubing: Material and ID / OD are different.
(8) Connector of Drain Tank: Material and OD to connect Tubing (7) is
different.
Drain Kit replacement Procedure
1 Select Instrument >> Instrument Control.
2 Select Maintenance >> Sample Introduction.
3 Set Nebulizer Pump at 0.1 rps and pump out the solvent in the drain line.
4 Remove Spray Chamber.
5 Remove the drain tubing from the Spray Chamber.
6 Replace the existing bushing in the connector with bushing (2) for
Organic Solvents.
7 Replace all of tubing, peristaltic Pump tubing and connectors between
Spray Chamber connector (1) to Drain tank (8) for Organic solvents.
8 Connect Tubing (7) to the Drain Tank for Organic Solvents.
9 Connect the Drain Kit to the Spray Chamber.
No in Drain Description
1 Connector for Spray Chamber Drainage
2 Bushing (Per flour)
3 Tubing (Polyethylene) ID 4.0mm OD 5.6mm length 35cm
7 Tubing (Polyethylene) ID 4.0mm OD 5.6mm length 120cm
4,6 Connector OD to connect (3) / (7) is 3.5mm
5 Peristaltic Pump Tubing (PharMed)
8 Drain Tank (with Polypropylene Connector)