6-14 Agilent 7500 Series ICP-MS Hardware Manual
6 Technical Information
such as the position of the mach disc, that are affected by the plasma
temperature; a lighter mass penetrates through the skimmer cone easier than
a heavier mass.
Figure 131 Distribution of Ion Energy
Extraction Lens
This lens assembly is placed just behind the skimmer cone and mounted on
the skimmer base. Its purpose is to extract ions passed through the skimmer
cone and to accelerate them to the Einzel lens. Two lenses are driven at
negative voltage. After the skimmer cone, there are positive ions, electrons,
and neutral atoms. When the positive ions come close to the extraction lenses,
the ions are accelerated. On the other hand, electrons are repelled by the
extraction lenses and go to ground via the skimmer cone and the chamber
wall.
Since the extraction lenses are placed very close to the plasma, they can
become contaminated. When the signal response in the tuning mode is slow
after changing the lens voltage, the lens may have become dirty and cleaning
will be required.
Low
Ion Energy
High
Ion Lens Setting
Sensitivity
Li
Y
T1