5-52 Agilent 7500 Series ICP-MS Hardware Manual
5 Troubleshooting
Items to Be Checked in Analysis Mode
• Check to make sure that the torch is positioned correctly.
Using Autotune, adjust the torch vertical/horizontal position
• Check to make sure that the mass axis is adjusted correctly.
Using Autotune, adjust the Resolution/Axis.
• Execute the EM voltage adjustment.
Using the Autotune, adjust the EM/discriminator.
• Check to make sure that the lens is responding.
Check to make sure that the signal changes when the lens voltage is varied.
• Check to make sure that sensitivity is acceptable in no gas mode.
If the sensitivity is acceptable in no gas mode and it is too low in He or H
2
mode, the problem may be caused by the purity level of the reaction gas.
• Check the sensitivity with mass 76 in no gas mode.
cps
• Set the tuning parameter as they were when the sensitivity was normal;
then re- check the sensitivity.