5-52 Agilent 7500 Series ICP-MS Hardware Manual
5 Troubleshooting
Items to Be Checked in Analysis Mode
Check to make sure that the torch is positioned correctly.
Using Autotune, adjust the torch vertical/horizontal position
Check to make sure that the mass axis is adjusted correctly.
Using Autotune, adjust the Resolution/Axis.
Execute the EM voltage adjustment.
Using the Autotune, adjust the EM/discriminator.
Check to make sure that the lens is responding.
Check to make sure that the signal changes when the lens voltage is varied.
Check to make sure that sensitivity is acceptable in no gas mode.
If the sensitivity is acceptable in no gas mode and it is too low in He or H
2
mode, the problem may be caused by the purity level of the reaction gas.
Check the sensitivity with mass 76 in no gas mode.
cps
Set the tuning parameter as they were when the sensitivity was normal;
then re- check the sensitivity.