Hardware manual
5-50 Agilent 7500 Series ICP-MS Hardware Manual
5 Troubleshooting
2606 Warning: Z- axis Torch position is outside the warning limit.
The Z- axis torch position moved beyond the warning limit.
2607 Warning: Y- Axis movement disabled due to partially closed top
cover.
The torch box tried to move, but failed because the small cover was not opened
fully.
Try it again after opening the small cover fully. There may be a fault in the
small cover Full Open Sensor or a fault in the System I/O board.
2608 Warning: Torch position initialization failed.
The torch failed to initialize its position. The torch cannot be controlled.
The XYZ stage was not managed to control properly. The following causes are
inferred.
• The connector of the XYZMN stage came off.
• Faulty XYZMN stage.
• Faulty XYZMN board.
2609 Warning: Penning gauge (AN) did not strike the first time. Will retry
within 120 seconds.
The Penning gauge failed to strike the first time. Retrying.
2610 Warning: Torch position hasn't initialized.
The Torch position (XYZ stage) is not initialized. The Torch position may be
incorrect.
Initialize the XYZ stage.
3000 The SC2 firmware has been successfully downloaded.
The firmware of the main CPU has finished being downloaded.
3001 The Z80 firmware has been successfully downloaded.
The firmware of the sub- CPU has finished being downloaded.
3002 Plasma OFF emergency switch was pushed.
The Plasma was extinguished urgently because the plasma OFF emergency
switch was pushed.
3003 Shutdown emergency switch was pushed.
The instrument starts the Shutdown sequence because the Shutdown
emergency switch was pushed. MassHunter Workstation (or ChemStation)
cannot control the instrument.
NOTE: After finishing the Shutdown emergency sequence, the MassHunter
Workstation (or ChemStation) cannot control the instrument unless the
instrument power is switched OFF. Check that the status lamp (LED) turns off
and switch off the power.