Hardware manual
5-44 Agilent 7500 Series ICP-MS Hardware Manual
5 Troubleshooting
1521 Error: System I/O Board controller communication Error.
The communication between the boards cannot be performed properly.
Check that all the communication lines and the power line are connected
properly. Faulty communication lines or Introduction board is inferred as the
cause.
1522 Error: QP Driver Board controller communication Error.
The communication between the boards cannot be performed properly.
Check that all the communication lines and the power line are connected
properly. Faulty communication lines or Introduction board is inferred as the
cause.
1523 Error: RF Driver Board controller communication Error.
The communication between the boards cannot be performed properly.
Check that all the communication lines and the power line are connected
properly. Faulty communication lines or Introduction board is inferred as the
cause.
1524 Error: HV Driver Board controller communication Error.
The communication between the boards cannot be performed properly.
Check that all the communication lines and the power line are connected
properly. Faulty communication lines or Introduction board is inferred as the
cause.
1525 Error: OctP Driver Board controller communication Error.
The communication between the boards cannot be performed properly.
Check that all the communication lines and the power line are connected
properly. Faulty communication lines or Introduction board is inferred as the
cause.
1526 Error: Reaction Cell Board controller communication Error.
The communication between the boards cannot be performed properly.
Check that all the communication lines and the power line are connected
properly. Faulty communication lines or Introduction board is inferred as the
cause.